Delays in the detection of hypoxemia due to site of pulse oximetry probe placement.
Journal of Clinical Anesthesia 1999 March
STUDY OBJECTIVES: To determine if there were any differences in the time to detect hypoxemia related to the site of peripheral pulse oximetry (ear, hand, and foot) during the rapid induction of hypoxemia in healthy volunteers.
DESIGN: Repeated-measures, longitudinal, observational study.
SETTING: Anesthesia clinical research area of the Department of Anesthesiology.
PATIENTS: 13 healthy volunteers, aged 18 to 44 years.
INTERVENTIONS: Nellcor N-200 (Nellcor, Inc., Pleasanton, CA) oximeter probes were placed at the ear, hand, and foot. All units were turned on simultaneously with averaging times set for 5 seconds and signals sampled at 2 Hz. A computer-controlled anesthesia circuit was employed to induce mild hypercapnia and hyperoxia (end-tidal gas partial pressures: PETCO2 = 42 +/- 2 mmHg and PETO2 = 130 mmHg) for 5 minutes. PETO2 was then decreased to 45 +/- 2 mmHg over 60 seconds and held at that value for 5 minutes.
MEASUREMENTS AND MAIN RESULTS: The mean differences in time (sec) for pulse oximeters to detect hypoxemia (read less than 90%) between probe sites were determined and compared. The following mean differences in time (sec) for pulse oximeters to detect hypoxemia (read less than 90%) between probe sites were found: ear-hand = 6; hand-foot = 57; ear-foot = 63. Paired t-tests revealed statistically significant mean time delay differences of 51 seconds (p < 0.005) and 57 seconds (p < 0.005) for ear-hand versus hand-foot and for ear-hand versus ear-foot, respectively.
CONCLUSIONS: In healthy volunteers, significant delays in the detection of acute hypoxemia by pulse oximetry occur when pulse oximeters are placed at the toe as compared with probes at either the ear or hand.
DESIGN: Repeated-measures, longitudinal, observational study.
SETTING: Anesthesia clinical research area of the Department of Anesthesiology.
PATIENTS: 13 healthy volunteers, aged 18 to 44 years.
INTERVENTIONS: Nellcor N-200 (Nellcor, Inc., Pleasanton, CA) oximeter probes were placed at the ear, hand, and foot. All units were turned on simultaneously with averaging times set for 5 seconds and signals sampled at 2 Hz. A computer-controlled anesthesia circuit was employed to induce mild hypercapnia and hyperoxia (end-tidal gas partial pressures: PETCO2 = 42 +/- 2 mmHg and PETO2 = 130 mmHg) for 5 minutes. PETO2 was then decreased to 45 +/- 2 mmHg over 60 seconds and held at that value for 5 minutes.
MEASUREMENTS AND MAIN RESULTS: The mean differences in time (sec) for pulse oximeters to detect hypoxemia (read less than 90%) between probe sites were determined and compared. The following mean differences in time (sec) for pulse oximeters to detect hypoxemia (read less than 90%) between probe sites were found: ear-hand = 6; hand-foot = 57; ear-foot = 63. Paired t-tests revealed statistically significant mean time delay differences of 51 seconds (p < 0.005) and 57 seconds (p < 0.005) for ear-hand versus hand-foot and for ear-hand versus ear-foot, respectively.
CONCLUSIONS: In healthy volunteers, significant delays in the detection of acute hypoxemia by pulse oximetry occur when pulse oximeters are placed at the toe as compared with probes at either the ear or hand.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app