Intraocular pressure measurement in patients with instrastromal corneal ring segments.
Journal of Refractive Surgery 1999 July
PURPOSE: To study the measurement of intraocular pressure after implantation of Intacs (ICRS) intrastromal corneal ring segment, a device that is positioned circumferentially in the peripheral corneal stroma to correct myopia. The device changes the corneal curvature by shortening arc length. Since the ring segments are made of polymethylmethacrylate, this may cause localized changes in corneal elasticity so intraocular pressure measurement may be affected.
METHODS: We measured the intraocular pressure of 12 eyes in which the ICRS had been in place longer than 6 months. We used Goldmann applanation and Tono-Pen tonometers over the central corneal and the paracentral corneal areas. We also measured the intraocular pressure with the Tono-Pen applanated directly over the intrastromal corneal ring segments.
RESULTS: The resulting intraocular pressure measurements were similar for the Tono-Pen tonometer readings over the central cornea, paracentral cornea, and the Goldmann applanation tonometer readings over the central cornea (P < .01). Our measurements using the Goldmann applanation tonometer on the paracentral corneal area showed artificially elevated intraocular pressure in the 40 to 60 mmHg range. We were not able to obtain consistent results when we measured the intraocular pressure using the Tono-Pen on the corneal area directly overlying the intrastromal corneal ring segment implants.
CONCLUSION: Consistent intraocular pressure measurements on eyes with the ICRS can be obtained with the Goldmann applanation tonometer over the central corneal area or with the Tono-Pen tonometer over the central or paracentral corneal areas.
METHODS: We measured the intraocular pressure of 12 eyes in which the ICRS had been in place longer than 6 months. We used Goldmann applanation and Tono-Pen tonometers over the central corneal and the paracentral corneal areas. We also measured the intraocular pressure with the Tono-Pen applanated directly over the intrastromal corneal ring segments.
RESULTS: The resulting intraocular pressure measurements were similar for the Tono-Pen tonometer readings over the central cornea, paracentral cornea, and the Goldmann applanation tonometer readings over the central cornea (P < .01). Our measurements using the Goldmann applanation tonometer on the paracentral corneal area showed artificially elevated intraocular pressure in the 40 to 60 mmHg range. We were not able to obtain consistent results when we measured the intraocular pressure using the Tono-Pen on the corneal area directly overlying the intrastromal corneal ring segment implants.
CONCLUSION: Consistent intraocular pressure measurements on eyes with the ICRS can be obtained with the Goldmann applanation tonometer over the central corneal area or with the Tono-Pen tonometer over the central or paracentral corneal areas.
Full text links
Related Resources
Trending Papers
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Prevention and treatment of ischaemic and haemorrhagic stroke in people with diabetes mellitus: a focus on glucose control and comorbidities.Diabetologia 2024 April 17
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Eosinophilic Esophagitis: Clinical Pearls for Primary Care Providers and Gastroenterologists.Mayo Clinic Proceedings 2024 April
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app