Influence of image acquisition parameters on CT artifacts and polyp depiction in spiral CT colonography: in vitro evaluation.
Radiology 2000 October
PURPOSE: To quantify the effects of spiral computed tomographic (CT) acquisition parameters on the magnitude of three-dimensional (3D) rippling artifacts and polyp depiction.
MATERIALS AND METHODS: An in vitro colon phantom was constructed with air-filled acrylic cylinders that contained synthetic polyps of 3-13 mm. The phantom was submerged in fluid and positioned at four angles of inclination relative to the z axis. Image data were acquired at collimation and pitch combinations of 3 mm and 1.67 and 5 mm and 1.6, respectively. Rippling artifacts were quantified by measuring the longitudinal variation of in-plane phantom edge width, and the influence of these artifacts on the depiction of pedunculated and sessile polyps was assessed qualitatively.
RESULTS: The in-plane magnitude of the rippling artifact was a function of the angle of inclination relative to the longitudinal axis and the table increment. The through-plane periodicity of the artifact was equal to one-half the table increment.
CONCLUSION: The table increment and angle of inclination of the surface of the object relative to the z axis determine the periodicity and magnitude of the rippling artifact at 3D spiral CT colonography. Although the depiction of small pedunculated polyps was not compromised, some sessile polyps were degraded by the artifact.
MATERIALS AND METHODS: An in vitro colon phantom was constructed with air-filled acrylic cylinders that contained synthetic polyps of 3-13 mm. The phantom was submerged in fluid and positioned at four angles of inclination relative to the z axis. Image data were acquired at collimation and pitch combinations of 3 mm and 1.67 and 5 mm and 1.6, respectively. Rippling artifacts were quantified by measuring the longitudinal variation of in-plane phantom edge width, and the influence of these artifacts on the depiction of pedunculated and sessile polyps was assessed qualitatively.
RESULTS: The in-plane magnitude of the rippling artifact was a function of the angle of inclination relative to the longitudinal axis and the table increment. The through-plane periodicity of the artifact was equal to one-half the table increment.
CONCLUSION: The table increment and angle of inclination of the surface of the object relative to the z axis determine the periodicity and magnitude of the rippling artifact at 3D spiral CT colonography. Although the depiction of small pedunculated polyps was not compromised, some sessile polyps were degraded by the artifact.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app