Comparison of commercially available (125)I-based RIA methods for the determination of circulating 25-hydroxyvitamin D.
Clinical Chemistry 2000 October
BACKGROUND: Measurement of circulating 25-hydroxyvitamin D [25(OH)D] is important in the management of metabolic bone disease. The aim of this study was to compare two widely used methods for the quantification of circulating 25(OH)D with attention to their abilities to measure 25-hydroxylated ergocalciferol (vitamin D(2)) [25(OH)D(2)] and cholecalciferol (vitamin D(3)) [25(OH)D(3)].
METHODS: We used two commercially available, Food and Drug Administration-approved, radioiodine ((125)I)-based RIA kits for the detection of 25(OH)D (DiaSorin, Stillwater, MN and IDS Ltd, Tyne and Wear, United Kingdom). These methods were tested for general assay performance, including antibody specificity. Results were compared with those of an HPLC-based direct ultraviolet detection method.
RESULTS: Within- and between-run CVs were </=10%. Both methods quantitatively recovered 25(OH)D(3) added to serum, but only the DiaSorin kit quantitatively recovered 25(OH)D(2). The primary antibody in the IDS kit had unequal reactivities with pure 25(OH)D(2) and 25(OH)D(3), whereas the DiaSorin primary antibody reacted with them equally. In 50 patient samples assayed by HPLC, the IDS method, but not the DiaSorin method, underestimated total circulating 25(OH)D when significant circulating 25(OH)D(2) was present in patient samples.
CONCLUSIONS: Some immunoassays may underestimate total 25(OH)D when 25(OH)D(2) constitutes an appreciable part of the total.
METHODS: We used two commercially available, Food and Drug Administration-approved, radioiodine ((125)I)-based RIA kits for the detection of 25(OH)D (DiaSorin, Stillwater, MN and IDS Ltd, Tyne and Wear, United Kingdom). These methods were tested for general assay performance, including antibody specificity. Results were compared with those of an HPLC-based direct ultraviolet detection method.
RESULTS: Within- and between-run CVs were </=10%. Both methods quantitatively recovered 25(OH)D(3) added to serum, but only the DiaSorin kit quantitatively recovered 25(OH)D(2). The primary antibody in the IDS kit had unequal reactivities with pure 25(OH)D(2) and 25(OH)D(3), whereas the DiaSorin primary antibody reacted with them equally. In 50 patient samples assayed by HPLC, the IDS method, but not the DiaSorin method, underestimated total circulating 25(OH)D when significant circulating 25(OH)D(2) was present in patient samples.
CONCLUSIONS: Some immunoassays may underestimate total 25(OH)D when 25(OH)D(2) constitutes an appreciable part of the total.
Full text links
Related Resources
Trending Papers
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Anti-Arrhythmic Effects of Heart Failure Guideline-Directed Medical Therapy and Their Role in the Prevention of Sudden Cardiac Death: From Beta-Blockers to Sodium-Glucose Cotransporter 2 Inhibitors and Beyond.Journal of Clinical Medicine 2024 Februrary 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app