Electrosurgery, pacemakers and ICDs: a survey of precautions and complications experienced by cutaneous surgeons.

BACKGROUND: Minimal information is available in the literature regarding the precautions implemented or complications experienced by cutaneous surgeons when electrosurgery is used in patients with pacemakers or implantable cardioverter-defibrillators (ICDs). The literature pertinent to dermatologists is primarily based on experiences of other surgical specialties and a generally recommended thorough perioperative evaluation.

OBJECTIVE: To determine what precautions are currently taken by cutaneous surgeons in patients with pacemakers or ICDs, and what types of complications have occurred due to electrosurgery in a dermatologic setting.

METHODS: In the winter of 2000, a survey was mailed to 419 U.S.-based members of the American College of Mohs Micrographic Surgery and Cutaneous Oncology (ACMMSCO).

RESULTS: A total of 166 (40%) surveys were returned. Routine precautions included utilizing short bursts of less than 5 seconds (71%), use of minimal power (61%), and avoiding use around the pacemaker or ICD (57%). The types of interference reported were skipped beats (eight patients), reprogramming of a pacemaker (six patients), firing of an ICD (four patients), asystole (three patients), bradycardia (two patients), depleted battery life of a pacemaker (one patient), and an unspecified tachyarrhythmia (one patient). Overall there was a low rate of complications (0.8 cases/100 years of surgical practice), with no reported significant morbidity or mortality. Bipolar forceps were utilized by 19% of respondents and were not associated with any incidences of interference.

CONCLUSIONS: Significant interference to pacemakers or ICDs rarely results from office-based electrosurgery. No clear community practice standards regarding precautions was evident from this survey. The use of bipolar forceps or true electrocautery are the better options when electrosurgey is required. These two modalities may necessitate fewer perioperative precautions than generally recommended, without compromising patient safety.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app