Immersion A-scan compared with partial coherence interferometry: outcomes analysis.
Journal of Cataract and Refractive Surgery 2002 Februrary
PURPOSE: To compare 2 methods of axial length measurement, immersion ultrasonography and partial coherence interferometry, and to elucidate surgical outcomes based on immersion measurements.
SETTING: Oregon Eye Institute, Eugene, Oregon, USA.
METHODS: Axial length measurements in 50 cataractous eyes were obtained by optical biometry (IOLMaster, Zeiss Humphrey Systems) and immersion ultrasound (Axis II, Quantel Medical), and the results were compared. Intraocular lens (IOL) power calculations in the same eyes after cataract extraction and posterior chamber IOL implantation were evaluated retrospectively based on the postoperative spherical equivalent prediction error.
RESULTS: Immersion ultrasonography and partial coherence interferometry measurements correlated in a highly positive manner (correlation coefficient = 0.996). Outcomes analysis demonstrated 92.0% of eyes were within +/-0.5 diopter of emmetropia based on immersion axial length measurements.
CONCLUSION: Immersion ultrasonography provided highly accurate axial length measurements and permitted highly accurate IOL power calculations.
SETTING: Oregon Eye Institute, Eugene, Oregon, USA.
METHODS: Axial length measurements in 50 cataractous eyes were obtained by optical biometry (IOLMaster, Zeiss Humphrey Systems) and immersion ultrasound (Axis II, Quantel Medical), and the results were compared. Intraocular lens (IOL) power calculations in the same eyes after cataract extraction and posterior chamber IOL implantation were evaluated retrospectively based on the postoperative spherical equivalent prediction error.
RESULTS: Immersion ultrasonography and partial coherence interferometry measurements correlated in a highly positive manner (correlation coefficient = 0.996). Outcomes analysis demonstrated 92.0% of eyes were within +/-0.5 diopter of emmetropia based on immersion axial length measurements.
CONCLUSION: Immersion ultrasonography provided highly accurate axial length measurements and permitted highly accurate IOL power calculations.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app