The Camino intracranial pressure sensor: is it optimal technology? An internal audit with a review of current intracranial pressure monitoring technologies.

Neurosurgery 2001 November
OBJECTIVE: To audit the reliability of the Camino intracranial pressure (ICP) sensor (Camino Laboratories, San Diego, CA) in our clinical practice as part of a continuing quality assurance program, and to assess its relative usefulness as compared with currently available ICP monitoring technologies that we reviewed.

DESIGN: Prospective audit of ICP device reliability and function in 50 patients with head injuries.

METHODS: Zero drift was recorded immediately after the ICP device was removed from the patient. Dynamic frequency response bench testing of each functioning catheter from 0 to 30 Hz and static calibration testing from 0 to 100 mmHg during environmental temperature variation from 22 to 40 degrees C were carried out.

RESULTS: Zero drift (range, -13 to 22 mmHg; median, -1 mmHg) was recorded immediately after the devices were removed from patients. Seventeen (50%) of the devices tested for zero drift had absolute drifts of at least 3 mmHg. There was no correlation between recorded zero drift and duration of monitoring (r = 0.154, P = 0.207). Five sensors (10% of those tested) failed during patient monitoring and were replaced. Static and dynamic calibration tests of the functioning sensors were within the manufacturer's specifications. However, the sensitivity of the devices to environmental temperature remains a problem.

CONCLUSION: The Camino ICP sensor remains one of the most popular ICP monitoring devices for use in patients with traumatic brain injuries. However, our recent in-house assessment demonstrated the robustness of the device to be less than adequate during routine practice. In this study, more than 50% exhibited zero drift greater than 3 mmHg, which is unacceptable in a catheter tip ICP monitoring device in which zero drift and calibration cannot be checked in vivo. A review of the literature revealed that other available ICP monitoring devices may prove to be more reliable and thus more appropriate for routine clinical measurement of ICP.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app