EVALUATION STUDY
JOURNAL ARTICLE
Evaluation of a new handheld biosensor for point-of-care testing of whole blood beta-hydroxybutyrate concentration.
Hong Kong Medical Journal 2002 June
OBJECTIVES: To evaluate performance characteristics of the newly available handheld combined glucose and ketone meter for beta-hydroxybutyrate measurement.
DESIGN: Laboratory method evaluation.
MAIN OUTCOME MEASURES: Accuracy of beta-hydroxybutyrate measurement and effect of acetoacetate interference at clinically important beta-hydroxybutyrate levels.
RESULTS: Deming regression analysis of beta-hydroxybutyrate measurements assessed by the ketone sensor and a laboratory enzymatic method revealed a coefficient of determination of 0.989 (P<0.001). Passing-Bablok regression analysis showed a linear relationship between the two methods, ie Y= -0.32+1.13X. The 95% confidence interval of the slope and y-intercept were: slope=1.13 (95% confidence interval, 1.04 to 1.22); intercept= -0.32 (95% confidence interval, -0.59 to -0.06). The Bland-Altman plot showed a small proportional bias between the two methods. The mean bias +/-2 standard deviations was between -0.53 and 0.67 mmol/L. Beta-hydroxybutyrate measurements made by the sensor were linear up to 6 mmol/L. Replicate analysis of two samples spiked with 3.6 mmol/L and 0.8 mmol/L of beta-hydroxybutyrate resulted in coefficients of variation of 3.3% and 13%, respectively. The presence of acetoacetate caused a negative interference in beta-hydroxybutyrate measurement. Beta-hydroxybutyrate recovery was 97.0% and 90.7% when the ketone body ratios were 6:1 and 3:1, respectively.
CONCLUSION: The analytical performance of the sensor, when operated according to manufacturer's instructions, could meet the needs of point-of-care beta-hydroxybutyrate measurement. Additional clinical studies are needed to assess the benefits of introducing such an assay in a clinical setting.
DESIGN: Laboratory method evaluation.
MAIN OUTCOME MEASURES: Accuracy of beta-hydroxybutyrate measurement and effect of acetoacetate interference at clinically important beta-hydroxybutyrate levels.
RESULTS: Deming regression analysis of beta-hydroxybutyrate measurements assessed by the ketone sensor and a laboratory enzymatic method revealed a coefficient of determination of 0.989 (P<0.001). Passing-Bablok regression analysis showed a linear relationship between the two methods, ie Y= -0.32+1.13X. The 95% confidence interval of the slope and y-intercept were: slope=1.13 (95% confidence interval, 1.04 to 1.22); intercept= -0.32 (95% confidence interval, -0.59 to -0.06). The Bland-Altman plot showed a small proportional bias between the two methods. The mean bias +/-2 standard deviations was between -0.53 and 0.67 mmol/L. Beta-hydroxybutyrate measurements made by the sensor were linear up to 6 mmol/L. Replicate analysis of two samples spiked with 3.6 mmol/L and 0.8 mmol/L of beta-hydroxybutyrate resulted in coefficients of variation of 3.3% and 13%, respectively. The presence of acetoacetate caused a negative interference in beta-hydroxybutyrate measurement. Beta-hydroxybutyrate recovery was 97.0% and 90.7% when the ketone body ratios were 6:1 and 3:1, respectively.
CONCLUSION: The analytical performance of the sensor, when operated according to manufacturer's instructions, could meet the needs of point-of-care beta-hydroxybutyrate measurement. Additional clinical studies are needed to assess the benefits of introducing such an assay in a clinical setting.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app