Clinical performance of an in-line, ex vivo point-of-care monitor: a multicenter study.
Clinical Chemistry 2002 November
BACKGROUND: The management of critically ill infants and neonates includes frequent determination of arterial blood gas, electrolyte, and hematocrit values. An objective of attached point-of-care patient monitoring is to provide clinically relevant data without the adverse consequences associated with serial phlebotomy.
METHODS: We prospectively determined the mean difference (and SD of the difference) from laboratory methods of an in-line, ex vivo monitor, the VIA LVM Blood Gas and Chemistry Monitoring System (VIA LVM Monitor; Metracor Technologies, Inc.), in 100 critically ill neonates and infants at seven children's hospitals. In doing so, we examined monitor stability with continuous use. In vivo patient test results from laboratory benchtop analyzers were compared with those from the VIA LVM Monitor on paired samples. In a separate in vitro comparison, benchtop analyzer and monitor test results were compared on whole-blood split samples.
RESULTS: A total of 1414 concurrent, paired-sample measurements were obtained. The mean differences (SD of differences) from laboratory methods and r values for the combined data for the VIA LVM Monitor from the seven sites were 0.001 (0.026) and 0.97 for pH, 0.7 (3.6) mmHg and 0.94 for PCO(2), 4.2 (9.6) mmHg and 0.98 for PO(2), 0.0 (2.9) mmol/L and 0.87 for sodium, 0.1 (0.2) mmol/L and 0.96 for potassium, and 0.3% (2.9%) and 0.90 for hematocrit. Performance results were similar among the study sites with increasing time of monitor use and between in vivo paired-sample and in vitro split-sample test results.
CONCLUSION: The VIA LVM Monitor can be used to assess critically ill neonates and infants.
METHODS: We prospectively determined the mean difference (and SD of the difference) from laboratory methods of an in-line, ex vivo monitor, the VIA LVM Blood Gas and Chemistry Monitoring System (VIA LVM Monitor; Metracor Technologies, Inc.), in 100 critically ill neonates and infants at seven children's hospitals. In doing so, we examined monitor stability with continuous use. In vivo patient test results from laboratory benchtop analyzers were compared with those from the VIA LVM Monitor on paired samples. In a separate in vitro comparison, benchtop analyzer and monitor test results were compared on whole-blood split samples.
RESULTS: A total of 1414 concurrent, paired-sample measurements were obtained. The mean differences (SD of differences) from laboratory methods and r values for the combined data for the VIA LVM Monitor from the seven sites were 0.001 (0.026) and 0.97 for pH, 0.7 (3.6) mmHg and 0.94 for PCO(2), 4.2 (9.6) mmHg and 0.98 for PO(2), 0.0 (2.9) mmol/L and 0.87 for sodium, 0.1 (0.2) mmol/L and 0.96 for potassium, and 0.3% (2.9%) and 0.90 for hematocrit. Performance results were similar among the study sites with increasing time of monitor use and between in vivo paired-sample and in vitro split-sample test results.
CONCLUSION: The VIA LVM Monitor can be used to assess critically ill neonates and infants.
Full text links
Related Resources
Trending Papers
Revascularization Strategy in Myocardial Infarction with Multivessel Disease.Journal of Clinical Medicine 2024 March 27
Intravenous infusion of dexmedetomidine during the surgery to prevent postoperative delirium and postoperative cognitive dysfunction undergoing non-cardiac surgery: a meta-analysis of randomized controlled trials.European Journal of Medical Research 2024 April 19
The Tricuspid Valve: A Review of Pathology, Imaging, and Current Treatment Options: A Scientific Statement From the American Heart Association.Circulation 2024 April 26
Consensus Statement on Vitamin D Status Assessment and Supplementation: Whys, Whens, and Hows.Endocrine Reviews 2024 April 28
Management of Diverticulitis: A Review.JAMA Surgery 2024 April 18
Interstitial Lung Disease: A Review.JAMA 2024 April 23
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app