Value of CT volume imaging for optimal placement of radiofrequency ablation probes in liver lesions.
Journal of Vascular and Interventional Radiology : JVIR 2002 November
PURPOSE: To determine the value of multiplanar and volume-rendered modes of analysis versus axial imaging for the optimal placement of probes in spiral image computed tomography (CT)-guided radiofrequency (RF) ablation of liver lesions.
MATERIALS AND METHODS: Thirty-two malignant hepatic lesions in 21 patients with hepatocellular carcinoma (n = 10) or metastases (n = 11) were treated with percutaneous RF ablation. After axial CT-guided placement (biopsy mode) of the probe in the tumor, a contrast-enhanced spiral image set was obtained and data were viewed in contiguous axial sections and the multiplanar and volume-rendered modes. The position of the probe was characterized as "central," "marginal," or "outside," corresponding with its position in the lesion. Outside positioning was corrected before ablation was performed and, in cases of "marginal" positioning, the probe was redirected after initial ablative therapy in the same session.
RESULTS: Multiplanar and volume-rendered analysis reclassified the needle position before 14 of 32 RF ablation procedures (44%; P =.0034, McNemar test). In 10 cases, probe positions considered "central" on biopsy-mode images were found to be "marginal" with multiplanar/volume-rendered analysis. In three cases, probe positions considered "marginal" on biopsy-mode images were found to be "outside" with multiplanar/volume-rendered analysis. In one lesion, multiplanar/volume-rendered analysis upgraded the probe position from "marginal" to "central." Comparison of multiplanar and volume-rendered analysis revealed no difference in probe positioning. The time requirements to obtain additional data were 159 seconds +/- 49 for multiplanar analysis and 434 seconds +/- 78 for volume-rendered analysis.
CONCLUSION: A contrast-enhanced spiral CT image data set with multiplanar reformations to verify an optimal probe location after axial CT-guided needle placement should be obtained before RF ablation of liver lesions.
MATERIALS AND METHODS: Thirty-two malignant hepatic lesions in 21 patients with hepatocellular carcinoma (n = 10) or metastases (n = 11) were treated with percutaneous RF ablation. After axial CT-guided placement (biopsy mode) of the probe in the tumor, a contrast-enhanced spiral image set was obtained and data were viewed in contiguous axial sections and the multiplanar and volume-rendered modes. The position of the probe was characterized as "central," "marginal," or "outside," corresponding with its position in the lesion. Outside positioning was corrected before ablation was performed and, in cases of "marginal" positioning, the probe was redirected after initial ablative therapy in the same session.
RESULTS: Multiplanar and volume-rendered analysis reclassified the needle position before 14 of 32 RF ablation procedures (44%; P =.0034, McNemar test). In 10 cases, probe positions considered "central" on biopsy-mode images were found to be "marginal" with multiplanar/volume-rendered analysis. In three cases, probe positions considered "marginal" on biopsy-mode images were found to be "outside" with multiplanar/volume-rendered analysis. In one lesion, multiplanar/volume-rendered analysis upgraded the probe position from "marginal" to "central." Comparison of multiplanar and volume-rendered analysis revealed no difference in probe positioning. The time requirements to obtain additional data were 159 seconds +/- 49 for multiplanar analysis and 434 seconds +/- 78 for volume-rendered analysis.
CONCLUSION: A contrast-enhanced spiral CT image data set with multiplanar reformations to verify an optimal probe location after axial CT-guided needle placement should be obtained before RF ablation of liver lesions.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
Perioperative echocardiographic strain analysis: what anesthesiologists should know.Canadian Journal of Anaesthesia 2024 April 11
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app