Activated clotting time systems vary in precision and bias and are not interchangeable when following heparin management protocols during cardiopulmonary bypass.

OBJECTIVE: Our aim was to test the hypothesis that new activated clotting time (ACT) technology, with modifications to instruments and reagents designed to detect earlier clot formation, would be associated with more precise but lower results. A secondary objective was to evaluate the potential impact of any change in ACT measurement on heparin requirements during cardiopulmonary bypass (CPB).

METHODS: We compared the precision of two newer ACT systems: Actalyke, Helena Laboratories, Beaumont, TX and Hemochron Response, International Technidyne Corporation, Edison, NJ and assessed their bias with reference to a standard ACT system (Hemochron 801, International Technidyne Corporation, Edison, NJ). Bland-Altman analysis was applied to 81 duplicate samples from 22 patients undergoing CPB or percutaneous coronary interventions (PCI), covering the full clinical range of ACT values. We also estimated the change in heparin dose required to use the Actalyke rather than the Hemochron 801 results, to achieve our target ACT for CPB (480 seconds), and used a mixed model to test for significance.

RESULTS: The precision of the Actalyke was superior to the Hemochron Response (mean difference of duplicates +/- 0.1% versus +/- 4.2%). There was no significant bias (p = 0.93) between the results from the standard analyzers (Hemochron 801 and Response), but the results from the modified system (Actalyke) were on average 18% lower than the Hemochron 801 (p < 0.0001). Estimated heparin requirements established that fifty percent of CPB patients would have required additional heparin (5000 to 17500 units), an average increase of 1060 units per patient (p = 0.05), if the Actalyke values were used to guide anticoagulation during CPB.

CONCLUSIONS: Our results support the hypothesis that the modified technology (Actalyke) is associated with more precise but lower ACT results. We estimated these lower values would lead to increased heparin dosing during CPB. The impact of this increase on bleeding after cardiac surgery with CPB is controversial and requires further study.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app