Accuracy of an electrochemical sensor for measuring capillary blood ketones by fingerstick samples during metabolic deterioration after continuous subcutaneous insulin infusion interruption in type 1 diabetic patients.
Diabetes Care 2003 April
OBJECTIVE: This study was designed to test the accuracy of capillary ketonemia for diagnosis of ketosis after interruption of insulin infusion.
RESEARCH DESIGN AND METHODS: A total of 18 patients with type 1 diabetes treated by external pump were studied during pump stop for 5 h. Plasma and capillary ketonemia and ketonuria were determined every hour from 7:00 A.M. (time 0 min = T0) to 12:00 P.M. (time 300 min = T300). Plasma beta-hydroxybutyrate (beta-OHB) levels were measured by an enzymatic end point spectrophotometric method, and capillary beta-OHB levels were measured by an electrochemical method (MediSense Optium meter). Ketonuria was measured by a semiquantitative test (Ketodiastix). Positive ketosis was defined by a value of >/=0.5 mmol/l for ketonemia and >/=4 mmol/l (moderate) for ketonuria.
RESULTS: After stopping the pump, concentrations of beta-OHB in both plasma and capillary blood increased significantly at time 60 min (T60) compared with T0 (P < 0.001), reaching maximum levels at T300 (1.30 +/- 0.49 and 1.23 +/- 0.78 mmol/l, respectively). Plasma and capillary beta-OHB values were highly correlated (r = 0.94, P < 0.0001). For diagnosis of ketosis, capillary ketonemia has a higher sensitivity and negative predictive value (80.4 and 82.5%, respectively) than ketonuria (63 and 71.8%, respectively). For plasma glucose levels >/=250 mg/dl, plasma and capillary ketonemia were found to be more frequently positive (85 and 78%, respectively) than ketonuria (59%) (P = 0.017). The time delay to diagnosis of ketosis was significantly higher for ketonuria than for plasma ketonemia (212 +/- 67 vs. 140 +/- 54 min, P = 0.0023), whereas no difference was noted between plasma and capillary ketonemia.
CONCLUSIONS: The frequency of screening for ketosis and the efficiency of detection of ketosis definitely may be improved by the use of capillary blood ketone determination in clinical practice.
RESEARCH DESIGN AND METHODS: A total of 18 patients with type 1 diabetes treated by external pump were studied during pump stop for 5 h. Plasma and capillary ketonemia and ketonuria were determined every hour from 7:00 A.M. (time 0 min = T0) to 12:00 P.M. (time 300 min = T300). Plasma beta-hydroxybutyrate (beta-OHB) levels were measured by an enzymatic end point spectrophotometric method, and capillary beta-OHB levels were measured by an electrochemical method (MediSense Optium meter). Ketonuria was measured by a semiquantitative test (Ketodiastix). Positive ketosis was defined by a value of >/=0.5 mmol/l for ketonemia and >/=4 mmol/l (moderate) for ketonuria.
RESULTS: After stopping the pump, concentrations of beta-OHB in both plasma and capillary blood increased significantly at time 60 min (T60) compared with T0 (P < 0.001), reaching maximum levels at T300 (1.30 +/- 0.49 and 1.23 +/- 0.78 mmol/l, respectively). Plasma and capillary beta-OHB values were highly correlated (r = 0.94, P < 0.0001). For diagnosis of ketosis, capillary ketonemia has a higher sensitivity and negative predictive value (80.4 and 82.5%, respectively) than ketonuria (63 and 71.8%, respectively). For plasma glucose levels >/=250 mg/dl, plasma and capillary ketonemia were found to be more frequently positive (85 and 78%, respectively) than ketonuria (59%) (P = 0.017). The time delay to diagnosis of ketosis was significantly higher for ketonuria than for plasma ketonemia (212 +/- 67 vs. 140 +/- 54 min, P = 0.0023), whereas no difference was noted between plasma and capillary ketonemia.
CONCLUSIONS: The frequency of screening for ketosis and the efficiency of detection of ketosis definitely may be improved by the use of capillary blood ketone determination in clinical practice.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app