Temperature monitoring during radiofrequency catheter ablation of accessory pathways.
Circulation 1992 November
BACKGROUND: Animal studies have suggested that the temperature of the electrode-tissue interface during radiofrequency catheter ablation accurately predicts lesion size. The purpose of the current study was to evaluate the utility of continuous temperature monitoring during radiofrequency catheter ablation in patients with Wolff-Parkinson-White syndrome.
METHODS AND RESULTS: Twenty patients with manifest preexcitation were included in the study. The ablation catheter was positioned on the ventricular side of the mitral annulus for left-sided accessory pathways and on the atrial side of the tricuspid annulus for right-sided and septal accessory pathways. A thermistor imbedded in the distal electrode of the ablation catheter allowed continuous temperature monitoring during each energy application. To define the relation between power and temperature, radiofrequency current was applied several times at each site using outputs of 20, 30, 40, and 50 W. The accessory pathways were successfully ablated in each of the 20 patients. Because of marked variability in the efficiency of heating between sites, power output did not predict temperature. However, at any given site, there was a positive dose-response relation between power and temperature. Radiofrequency energy applications on the atrial side of the tricuspid annulus produced lower temperatures than did applications on the ventricular side of the mitral annulus (49 +/- 7 versus 60 +/- 16 degrees C, p = 0.0001). Transient block in the accessory pathways occurred at a mean of 50 +/- 8 degrees C, whereas permanent block was seen at a mean of 62 +/- 15 degrees C (p = 0.0001). Less than half of the applications at outputs < or = 40 W produced temperatures adequate to interrupt accessory pathway conduction. An abrupt rise in impedance caused by coagulum formation occurred only at temperatures between 95 and 100 degrees C.
CONCLUSIONS: Temperature monitoring may facilitate radiofrequency catheter ablation of accessory pathways. By adjusting power output to ensure that adequate but not excessive temperatures have been achieved, a rise in impedance can be avoided and the total number of energy applications and procedure duration may be reduced.
METHODS AND RESULTS: Twenty patients with manifest preexcitation were included in the study. The ablation catheter was positioned on the ventricular side of the mitral annulus for left-sided accessory pathways and on the atrial side of the tricuspid annulus for right-sided and septal accessory pathways. A thermistor imbedded in the distal electrode of the ablation catheter allowed continuous temperature monitoring during each energy application. To define the relation between power and temperature, radiofrequency current was applied several times at each site using outputs of 20, 30, 40, and 50 W. The accessory pathways were successfully ablated in each of the 20 patients. Because of marked variability in the efficiency of heating between sites, power output did not predict temperature. However, at any given site, there was a positive dose-response relation between power and temperature. Radiofrequency energy applications on the atrial side of the tricuspid annulus produced lower temperatures than did applications on the ventricular side of the mitral annulus (49 +/- 7 versus 60 +/- 16 degrees C, p = 0.0001). Transient block in the accessory pathways occurred at a mean of 50 +/- 8 degrees C, whereas permanent block was seen at a mean of 62 +/- 15 degrees C (p = 0.0001). Less than half of the applications at outputs < or = 40 W produced temperatures adequate to interrupt accessory pathway conduction. An abrupt rise in impedance caused by coagulum formation occurred only at temperatures between 95 and 100 degrees C.
CONCLUSIONS: Temperature monitoring may facilitate radiofrequency catheter ablation of accessory pathways. By adjusting power output to ensure that adequate but not excessive temperatures have been achieved, a rise in impedance can be avoided and the total number of energy applications and procedure duration may be reduced.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app