Accuracy of urinalysis dipstick techniques in predicting significant proteinuria in pregnancy.

OBJECTIVE: To estimate the accuracy of point-of-care dipstick urinalysis in predicting significant proteinuria in pregnancy.

DATA SOURCES: Literature from 1970 to February 2002 was identified via 1). general bibliographic databases, that is, MEDLINE and EMBASE, 2). Cochrane Library and relevant specialist register of the Cochrane Collaboration, and 3). checking the reference lists of known primary and review articles.

METHODS OF STUDY SELECTION: Studies were selected if the accuracy of dipstick urinalysis techniques in predicting total protein excretion was estimated compared with a reference standard (laboratory estimation of protein excretion). The tests included visually read color-change dipsticks and automated dipstick urinalysis. Study selection, quality assessment, and data abstraction were performed independently and in duplicate.

TABULATION, INTEGRATION, AND RESULTS: Data from selected studies were abstracted as 2 x 2 tables comparing the test result with the reference standard. Test accuracy was expressed as likelihood ratios. Summary likelihood ratios were generated as measures of diagnostic accuracy to determine posttest probabilities. The electronic search produced 1543 citations. After independent review of published articles, a total of 34 articles was obtained for further scrutiny, and 7 studies were considered eligible for inclusion in the review. The 6 studies evaluating visual dipstick urinalysis produced a pooled positive likelihood ratio of 3.48 (95% confidence interval 1.66, 7.27) and a pooled negative likelihood ratio of 0.6 (95% confidence interval 0.45, 0.8) for predicting 300 mg/24-hour proteinuria at the 1+ or greater threshold.

CONCLUSION: The accuracy of dipstick urinalysis with a 1+ threshold in the prediction of significant proteinuria is poor and therefore of limited usefulness to the clinician. Accuracy may be improved at higher thresholds (greater than 1+ proteinuria), but available data are sparse and of poor methodological quality. Therefore, it is not possible to make meaningful inferences about accuracy at higher urine dipstick thresholds. There is an urgent need for research in this area of common obstetric practice.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app