Essure hysteroscopic sterilization: results based on utilizing a new coil catheter delivery system.
STUDY OBJECTIVE: To assess the safety and placement effectiveness of a new delivery catheter for the Essure micro-insert hysteroscopic sterilization system.
DESIGN: Prospective, multicenter, single-arm clinical study. (Canadian Task Force classification II-3).
SETTING: Hospital-based clinical research centers.
PATIENTS: One hundred two women of reproductive age and proven fertility.
INTERVENTION: A new coil catheter delivery system was used for micro-insert placement.
MEASUREMENTS AND MAIN RESULTS: The bilateral placement of micro-inserts using the coil catheter delivery system occurred in 100 women out of 102 attempts (98%). The two women who did not have successful micro-insert placement were subsequently shown to have proximal tubal stenotic disease.
CONCLUSION: These findings indicate that the introduction of the coil catheter delivery system for Essure hysteroscopic sterilization improves the micro-insert bilateral placement rate when compared with previous studies. It is hypothesized that the coil catheter has superior navigational properties, particularly within tortuous or narrowed tubal lumens.
DESIGN: Prospective, multicenter, single-arm clinical study. (Canadian Task Force classification II-3).
SETTING: Hospital-based clinical research centers.
PATIENTS: One hundred two women of reproductive age and proven fertility.
INTERVENTION: A new coil catheter delivery system was used for micro-insert placement.
MEASUREMENTS AND MAIN RESULTS: The bilateral placement of micro-inserts using the coil catheter delivery system occurred in 100 women out of 102 attempts (98%). The two women who did not have successful micro-insert placement were subsequently shown to have proximal tubal stenotic disease.
CONCLUSION: These findings indicate that the introduction of the coil catheter delivery system for Essure hysteroscopic sterilization improves the micro-insert bilateral placement rate when compared with previous studies. It is hypothesized that the coil catheter has superior navigational properties, particularly within tortuous or narrowed tubal lumens.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app