Comparative Study
Journal Article
Add like
Add dislike
Add to saved papers

False-negative results at multi-detector row CT colonography: multivariate analysis of causes for missed lesions.

Radiology 2005 May
PURPOSE: To determine causes of false-negative results at multi-detector row computed tomographic (CT) colonography and determine presumptive causes with logistic regression analysis.

MATERIALS AND METHODS: Institutional review board approval and informed consent were obtained. The study included 394 colonic segments in 31 men and 25 women at high risk for colorectal cancer (mean age +/- standard deviation, 60.2 years +/- 9.3 for men and 56.8 years +/- 13.3 for women). Multi-detector row CT colonography and colonoscopy (reference standard) were performed in a blinded manner, and the results were compared. CT colonographic findings were interpreted in consensus by two readers using a primary two-dimensional with three-dimensional problem-solving approach. Adequacy of colonic preparation and distention was recorded. Sensitivity and specificity were obtained with 95% confidence intervals (CIs). Lesions missed at CT colonography were retrospectively reassessed to identify why they were missed, and, if the causes were not apparent, logistic regression analysis was performed to determine the presumptive causes.

RESULTS: Colonic preparation and distention were optimal in 17 patients (30%) but suboptimal in 37 (66%) and poor enough to make the results nondiagnostic in two (4%). Twenty-nine of 63 lesions were missed at CT colonography. When all flat, sessile, and pedunculated lesions (n = 63) were included, sensitivities were 75% (nine of 12; 95% CI: 48%, 100%) for lesions 10 mm or larger and 79% (19 of 24; 95% CI: 65%, 93%) for those 6 mm or larger. When only sessile and pedunculated lesions (n = 60) were included, corresponding sensitivities were 100% (nine of nine; 73%, 100%) and 90% (19 of 21; 78%, 100%), respectively. All three missed lesions larger than 10 mm were flat, and all three flat lesions were missed. Two 3-mm high lesions, including one invasive adenocarcinoma, were misinterpreted as feces at blinded image review; one 1-mm high tubular adenoma with adenocarcinoma foci could not be visualized even in retrospect. Sessile or pedunculated polyps 5 mm or smaller were significantly more likely to be missed than those 6 mm or larger (adjusted odds ratio, 11.6; P = .027).

CONCLUSION: Aside from inadequate bowel preparation and/or distention, flat lesions and small polyps are the two main causes for missed lesions at multi-detector row CT colonography.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app