Ureteric access with flexible ureteroscopes: effect of the size of the ureteroscope.
BJU International 2005 May
OBJECTIVE: To determine the optimum size of a flexible ureteroscope (FU) to minimize the need for ureteric dilatation.
MATERIALS AND METHODS: FUs from several different manufacturers and with different shaft diameters were used for ureteroscopy in adults. An initial attempt was made to pass one of the larger diameter FUs; if this was unsuccessful, the smaller 7.5 F FU was used. The results were reviewed retrospectively.
RESULTS: Flexible ureteroscopy was used in 115 consecutive patients at two different institutions. The ability to pass the FUs with no formal dilatation was directly related to the FU diameter. The rate of failing to pass with no formal dilatation was 37% for the 9.0 F, 8.5% for the 8.6 F, 5% for the 8.4 F and 0.9% for the 7.4 F ureteroscope.
CONCLUSION: Although other factors influence the overall effectiveness of the FU, the ability to pass FUs is directly related to their outer diameter. When only the ease of introduction is considered, the ideal ureteroscope outer diameter is 7.4 F.
MATERIALS AND METHODS: FUs from several different manufacturers and with different shaft diameters were used for ureteroscopy in adults. An initial attempt was made to pass one of the larger diameter FUs; if this was unsuccessful, the smaller 7.5 F FU was used. The results were reviewed retrospectively.
RESULTS: Flexible ureteroscopy was used in 115 consecutive patients at two different institutions. The ability to pass the FUs with no formal dilatation was directly related to the FU diameter. The rate of failing to pass with no formal dilatation was 37% for the 9.0 F, 8.5% for the 8.6 F, 5% for the 8.4 F and 0.9% for the 7.4 F ureteroscope.
CONCLUSION: Although other factors influence the overall effectiveness of the FU, the ability to pass FUs is directly related to their outer diameter. When only the ease of introduction is considered, the ideal ureteroscope outer diameter is 7.4 F.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app