Confirmation of Essure microinsert tubal coil placement with conventional and volume-contrast imaging three-dimensional ultrasound.
Fertility and Sterility 2005 August
OBJECTIVE: To determine the accuracy of ultrasound in the assessment of proximal fallopian tube positioning of the Essure microinsert coil 3 months after postprocedure.
DESIGN: Prospective cohort study (Canadian Task Force classification II-2).
SETTING: Reproductive-age women in a tertiary care hospital.
PATIENT(S): Reproductive-age women presenting with a request for permanent contraception.
INTERVENTION(S): Hysteroscopic sterilization with the Essure microinsert coil and conventional or volume-contrast three-dimensional (3D) ultrasound imaging 3 months after the procedure.
MAIN OUTCOME MEASURE(S): Coil position on ultrasound.
RESULT(S): Forty-eight of the 50 patients had successful placement of the Essure coils, and three patients required a second attempt on one tube. Conventional or volume-contrast (3D) ultrasound showed proper positioning of the coils within the proximal fallopian tube in 42 women (84%); five women (10%) required hysterosalpingogram to show appropriate positioning. Two patients (4%) required laparoscopic tubal sterilization, and one patient (2%) was lost to follow-up.
CONCLUSION(S): Transvaginal ultrasound is an acceptable method of confirming proper placement of the Essure microinsert coil within the proximal fallopian tube 3 months after the procedure.
DESIGN: Prospective cohort study (Canadian Task Force classification II-2).
SETTING: Reproductive-age women in a tertiary care hospital.
PATIENT(S): Reproductive-age women presenting with a request for permanent contraception.
INTERVENTION(S): Hysteroscopic sterilization with the Essure microinsert coil and conventional or volume-contrast three-dimensional (3D) ultrasound imaging 3 months after the procedure.
MAIN OUTCOME MEASURE(S): Coil position on ultrasound.
RESULT(S): Forty-eight of the 50 patients had successful placement of the Essure coils, and three patients required a second attempt on one tube. Conventional or volume-contrast (3D) ultrasound showed proper positioning of the coils within the proximal fallopian tube in 42 women (84%); five women (10%) required hysterosalpingogram to show appropriate positioning. Two patients (4%) required laparoscopic tubal sterilization, and one patient (2%) was lost to follow-up.
CONCLUSION(S): Transvaginal ultrasound is an acceptable method of confirming proper placement of the Essure microinsert coil within the proximal fallopian tube 3 months after the procedure.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app