Low power ultrasound delivered through a PTCA-like guidewire: preclinical feasibility and safety of a novel technology for intracoronary thrombolysis.

BACKGROUND: Low power ultrasound delivered through an angioplasty-like guidewire may be effective for intracoronary thrombolysis. We evaluated the preclinical feasibility and safety of such wire.

METHODS AND RESULTS: In 15 anesthetized Yucatan minipigs, the ultrasonic wire was advanced percutaneously into all three coronaries. Each coronary was randomized to long activation (6 minutes), short activation (3 minutes), or control (3 minutes indwelling, no activation). The energy delivered was 0.14 +/- 0.01 W/cm of active length (20 kHz). No changes in heart rate, rhythm, or arterial pressure occurred during wire positioning or activation. Mean lumen diameter (MLD) by quantitative angiography was not significantly different pre- and postintervention (2.36 +/- 0.12 mm vs 2.36 +/- 0.11 mm for long activation, P = 0.96; 2.33 +/- 0.15 mm vs 2.34 +/- 0.14 mm for short activation, P = 0.54; 2.30 +/- 0.12 mm vs 2.33 +/- 0.12 mm for control, P = 0.21). There were no angiographic stenoses at 60 or 90 days follow-up. Compared with baseline, MLD at follow-up increased in all the three groups (2.40 +/- 0.13 mm vs 2.53 +/- 0.11 mm, P = 0.004 for long activation; 2.37 +/- 0.17 mm vs 2.52 +/- 0.14 mm, P = 0.023 for short activation; 2.20 +/- 0.12 mm vs 2.33 +/- 0.11 mm, P = 0.001 for the control group). By histology, there were no clinically significant pathologic changes in coronary morphology.

CONCLUSION: Use of a transverse cavitation therapeutic wire is feasible and well tolerated acutely in the normal porcine coronary. At 60 and 90 days, no angiographically apparent damage, no clinically significant pathologic changes, and no adverse events were seen. This technology may be safely used during percutaneous coronary intervention. Further studies are justified to evaluate its efficacy for intracoronary thrombus ablation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app