Validation a portable monitoring device for sleep apnea diagnosis in a population based cohort using synchronized home polysomnography.

Sleep 2006 March
SUBJECT OBJECTIVE: To assess the accuracy of a portable monitoring device based on peripheral arterial tonometry to diagnose obstructive sleep apnea (OSA). To propose a new standard for limited-channel device validation using synchronized polysomnography (PSG) home recordings and a population-based cohort.

DESIGN: Single-night, unattended PSG and Watch_PAT 100 (WP_100).

SETTING: Home environment.

PARTICIPANTS: Ninety-eight subjects (55 men; age, 60 +/- 7 year; body mass index, 28 +/- 4 kg/m2) consecutively recruited from the Skaraborg Hypertension and Diabetes Project.

MEASUREMENTS AND RESULTS: The WP_100 records peripheral arterial tone, heart rate, oxygen saturation and actigraphy for automatic analysis of respiratory disturbance index (RDI), apnea-hypopnea index (AHI), oxygen desaturation index (ODI), and sleep-wake state. The accuracy of WP_100 in RDI, AHI, ODI, and sleep-wake detection was assessed by comparison with data from simultaneous PSG recordings. The mean PSG-AHI in this population was 25.5 +/- 22.9 events per hour. The WP_100 RDI, AHI, and ODI correlated closely (0.88, 0.90, and 0.92; p < .0001, respectively) with the corresponding indexes obtained by PSG. The areas under the curve for the receiver-operator characteristic curves for WP_100 AHI and RDI were 0.93 and 0.90 for the PSG-AHI and RDI thresholds 10 and 20 (p < .0001, respectively). The agreement of the sleep-wake assessment based on 30-second bins between the 2 systems was 82 +/- 7%.

CONCLUSIONS: The WP_100 was reasonably accurate for unattended home diagnosis of OSA in a population sample not preselected for OSA symptoms. The current design, including simultaneous home PSG recordings in population-based cohorts, is proposed as a reasonable validation standard for assessment of simplified recording tools for OSA diagnosis.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app