Optimization of thermocautery in excisional dermatologic surgery.
Dermatologic Surgery : Official Publication for American Society for Dermatologic Surgery [et Al.] 2006 May
PURPOSE: Electrosurgery is routinely used in cutaneous surgery for hemostasis. Thermocautery can be used in patients with implantable cardiac devices. This technique relies on heat without electrical current passing through the patient to produce hemostasis. The temperatures and utility of a commercially available, adjustable thermocautery unit are examined.
METHODOLOGY: Tip temperature of the commercially available thermocautery unit was measured in air and tissue via a type E thermocouple (0.002 in. diameter) around the unit's tip. Time intervals of 20 to 30 seconds were recorded at device settings 1 to 9 in air and 3 to 8 on surgical patients (Institutional Review Board approval obtained).
RESULTS: In vitro analysis demonstrated predictable temperatures at increasing settings in air: 350 to 900 degrees C. Analysis in vivo during surgery demonstrated similar findings. Tissue contact decreased tip temperature by approximately 50% from in vitro values, and use in a bloody field caused a further decrease in the tip temperature.
CONCLUSION: The thermocautery unit examined is an effective and safe unit to achieve hemostasis. In addition, the temperature may be adjusted as opposed to hand-held units that operate at in vitro temperatures exceeding 1,400 degrees F. Hemostasis at approximately 100 to 400 degrees C provided optimal hemostasis.
METHODOLOGY: Tip temperature of the commercially available thermocautery unit was measured in air and tissue via a type E thermocouple (0.002 in. diameter) around the unit's tip. Time intervals of 20 to 30 seconds were recorded at device settings 1 to 9 in air and 3 to 8 on surgical patients (Institutional Review Board approval obtained).
RESULTS: In vitro analysis demonstrated predictable temperatures at increasing settings in air: 350 to 900 degrees C. Analysis in vivo during surgery demonstrated similar findings. Tissue contact decreased tip temperature by approximately 50% from in vitro values, and use in a bloody field caused a further decrease in the tip temperature.
CONCLUSION: The thermocautery unit examined is an effective and safe unit to achieve hemostasis. In addition, the temperature may be adjusted as opposed to hand-held units that operate at in vitro temperatures exceeding 1,400 degrees F. Hemostasis at approximately 100 to 400 degrees C provided optimal hemostasis.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app