Randomized, double-blind, multicenter study of the Endeavor zotarolimus-eluting phosphorylcholine-encapsulated stent for treatment of native coronary artery lesions: clinical and angiographic results of the ENDEAVOR II trial.

Circulation 2006 August 23
BACKGROUND: The use of the Endeavor stent might reduce restenosis and stent thrombosis at 9 months.

METHODS AND RESULTS: Patients (n = 1197) treated for single coronary artery stenosis were enrolled in a prospective, randomized, double-blind study and randomly assigned to receive the Endeavor zotarolimus-eluting phosphorylcholine polymer-coated stent (n = 598) or the same bare metal stent but without the drug or the polymer coating (n = 599). The 2 groups were well matched in baseline characteristics. Diabetes was present in 20.1% of patients; the mean reference vessel diameter was 2.75 mm; and the mean lesion length was 14.2 mm. The primary end point of target vessel failure at 9 months was reduced from 15.1% with the bare metal stent to 7.9% with the Endeavor (P = 0.0001), and the rate of major adverse cardiac events was reduced from 14.4% with the bare metal stent to 7.3% with the Endeavor (P = 0.0001). Target lesion revascularization was 4.6% with Endeavor compared with 11.8% with the bare metal stent (P = 0.0001). The rate of stent thrombosis was 0.5% with the Endeavor, which was not significantly different from 1.2% with the bare metal stent. In 531 patients submitted to angiographic follow-up, late loss was reduced from 1.03 +/- 0.58 to 0.61 +/- 0.46 (P < 0.001) in stent and from 0.72 +/- 0.61 to 0.36 +/- 0.46 (P < 0.001) in segment. The rate of in-segment restenosis was reduced from 35.0% to 13.2% with Endeavor (P < 0.0001). There was no excessive edge stenosis, aneurysm formation, or late acquired malapposition by intravascular ultrasound imaging. Differences in clinical outcome were maintained at 12 and 24 months (P < 0.0001).

CONCLUSIONS: Compared with bare metal stents, the Endeavor stent is safe and reduces the rates of clinical and angiographic restenosis at 9, 12, and 24 months.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app