A comparison of the use of Trachlight and Eschmann multiple-use introducer in simulated difficult intubation.
European Journal of Anaesthesiology 2007 January
BACKGROUND AND OBJECTIVE: The Eschmann multiple-use introducer is widely used in the management of difficult intubations. Transillumination of the neck is less commonly used. We conducted a randomized crossover study comparing the Trachlight lightwand and Eschmann multiple-use introducer in simulated difficult intubation.
METHODS: Sixty-four healthy patients were studied using a standard anaesthetic and full muscle relaxation assessed by train of four. A Macintosh laryngoscope was then inserted and then lowered to simulate a Grade 3 view. Tracheal placement was attempted with both Trachlight lightwand and Eschmann multiple-use introducer in a randomized order. Anaesthetists placing the devices had extensive experience with the Eschmann multiple-use introducer, but only 15 previous uses of the Trachlight. Success rates and time for tracheal placement were recorded.
RESULTS: The Eschmann multiple-use introducer and Trachlight were successfully placed in 96.8% and 93.7%, respectively (n.s.). Mean (SD) time to intubation for Eschmann multiple-use introducer and Trachlight were 15(6) and *21(13), respectively (*P < 0.001).
CONCLUSION: The Trachlight is a potentially useful alternative to the Eschmann multiple-use introducer in difficult intubation.
METHODS: Sixty-four healthy patients were studied using a standard anaesthetic and full muscle relaxation assessed by train of four. A Macintosh laryngoscope was then inserted and then lowered to simulate a Grade 3 view. Tracheal placement was attempted with both Trachlight lightwand and Eschmann multiple-use introducer in a randomized order. Anaesthetists placing the devices had extensive experience with the Eschmann multiple-use introducer, but only 15 previous uses of the Trachlight. Success rates and time for tracheal placement were recorded.
RESULTS: The Eschmann multiple-use introducer and Trachlight were successfully placed in 96.8% and 93.7%, respectively (n.s.). Mean (SD) time to intubation for Eschmann multiple-use introducer and Trachlight were 15(6) and *21(13), respectively (*P < 0.001).
CONCLUSION: The Trachlight is a potentially useful alternative to the Eschmann multiple-use introducer in difficult intubation.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app