Adjustable vs set-pressure valves decrease the risk of proximal shunt obstruction in the treatment of pediatric hydrocephalus.

INTRODUCTION: The use of programmable shunt valves has increased dramatically in the practice of pediatric hydrocephalus. Despite theoretical advantages, it remains unclear if the use of programmable vs set-pressure valves affects shunt outcome.

MATERIALS AND METHODS: The clinical and radiological records of all pediatric patients undergoing ventriculoperitoneal (VP), ventriculopleural (VPl), and ventriculoatrial (VA) shunt surgery from 2001 to 2004 at an academic institution were reviewed. The association of programmable vs set-pressure valves with subsequent shunt revision was assessed by Kaplan-Meier shunt survival plots and log-rank analysis.

RESULTS: A total of 279 VP, VPl, and VA shunt surgeries were performed on patients with median (interquartile range) age of 4 (1-14) years (161 male, 118 female; 158 communicating, 122 obstructive hydrocephalus). Programmable valves were used in 76 (27%) cases and set-pressure valves in 203 (73%). At mean+/-SD follow-up of 17 +/- 13 months, programmable vs set-pressure valves were associated with reduced risk of both overall shunt revision [26 (35%) vs 109 (54%); relative risk (RR) (95% CI); 0.61 (0.41-0.91), p = 0.016] and proximal obstruction [9 (12%) vs 58 (28%); RR (95% CI); 0.39 (0.27-0.80), p = 0.006]. There was no difference in distal obstruction [3 (4%) vs 11 (5%) cases], infection [6 (8%) vs 12 (6%) cases], valve obstruction [0 (0%) vs 4 (2%)], or shunt disconnection [2 (3%) vs 1 (1%)] between adjustable and set-pressure valves, respectively.

CONCLUSION: In our experience, the use of programmable vales was associated with a decreased risk of proximal shunt obstruction and shunt revision. Programmable valves may be preferred in patients frequently experiencing proximal shunt failure. A prospective, controlled study is warranted to evaluate the potential value of adjustable vs set-pressure valve systems.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app