Coronary stent assessability by 64 slice multi-detector computed tomography.
Catheterization and Cardiovascular Interventions 2007 June 2
BACKGROUND: We evaluated the assessability of contemporary stent platforms by 64-slice multi-detector computed tomography (MDCT).
METHODS: Patients undergoing coronary stenting were included in a prospective protocol of MDCT imaging within 48 hr of stent implantation. MDCT data were acquired using a "Sensation 64" MDCT scanner (Siemens Medical Solutions, Forchheim, Germany). Stent assessability was assessed by two independent blinded observers and disagreement was resolved by a third observer. Assessability was defined at visualization of the in-stent lumen without influence of partial volume effects, beam hardening, motion, calcification, or contrast to noise limitations.
RESULTS: Fifty four stents (Cypher n = 25, Vision/Minivision n = 19, Taxus Express n = 8, Liberte n = 1, Driver n = 1) in 44 patients were included in the study. The two independent observers classified 30 of 54 stents (56%) as assessable. Interobserver reproducibility was good with kappa = 0.66. Stent size was the most important determinant of assessability. Consistently assessable stents were 3.0 mm or larger (85%), whereas those under 3 mm were mostly nonassessable (26%).
CONCLUSIONS: Contemporary stent designs evaluated on a 64-slice MDCT scanner showed artifact free assessability only in larger stents. Increase in spatial resolution of MDCT scanners or modifications in stent design will be necessary to noninvasive evaluate stents <3 mm in diameter, where in-stent restenosis is more frequent.
METHODS: Patients undergoing coronary stenting were included in a prospective protocol of MDCT imaging within 48 hr of stent implantation. MDCT data were acquired using a "Sensation 64" MDCT scanner (Siemens Medical Solutions, Forchheim, Germany). Stent assessability was assessed by two independent blinded observers and disagreement was resolved by a third observer. Assessability was defined at visualization of the in-stent lumen without influence of partial volume effects, beam hardening, motion, calcification, or contrast to noise limitations.
RESULTS: Fifty four stents (Cypher n = 25, Vision/Minivision n = 19, Taxus Express n = 8, Liberte n = 1, Driver n = 1) in 44 patients were included in the study. The two independent observers classified 30 of 54 stents (56%) as assessable. Interobserver reproducibility was good with kappa = 0.66. Stent size was the most important determinant of assessability. Consistently assessable stents were 3.0 mm or larger (85%), whereas those under 3 mm were mostly nonassessable (26%).
CONCLUSIONS: Contemporary stent designs evaluated on a 64-slice MDCT scanner showed artifact free assessability only in larger stents. Increase in spatial resolution of MDCT scanners or modifications in stent design will be necessary to noninvasive evaluate stents <3 mm in diameter, where in-stent restenosis is more frequent.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app