Comparison of forehead and digit oximetry in surgical/trauma patients at risk for decreased peripheral perfusion.

OBJECTIVE: Measurement of pulse oximetry (Spo(2)) is often impaired in critically ill patients. Forehead reflectance oximetry, the Max-Fast (Nellcor, Pleasanton, CA), may be less susceptible to poor tissue perfusion and could improve accuracy of oxygen saturation measurement. The objective of this study was to evaluate the use of forehead oximetry measures in critically ill surgical/trauma patients.

METHODS: A prospective interventional study of 30 critically ill surgical/trauma patients at risk for decreased peripheral perfusion, as evidenced by need for vasopressor support (24 patients), transfusion of more than 6 unit packed cells in 24 hours (two patients), or an inability to obtain consistent saturation from a digit sensor (four patients), compared forehead and digit-based oximeter Spo(2) readings with co-oximetry (Sao(2)) measurements from arterial blood samples. Sao(2) values were converted to functional oxygen saturation (SO(2)) measurements for the final comparison. Patients were fitted with forehead (Nellcor Max-Fast) and digit (Nellcor Max A; digit 1) sensors connected to Nellcor OxiMax N-595 oximeters and a digit sensor (Nellcor Max A; digit 2) connected to a multiparameter monitor (Philips CMS [Andover, MA]). Three measurements of Sao(2) were obtained from each subject over a 24-hour time period, and simultaneous measurements of Spo(2) were recorded from the three monitors.

RESULTS: The three Spo(2) measurements (forehead, digit 1, and digit 2) were compared with SO(2) values using the Bland-Altman method to assess agreement. Forehead measurements demonstrated a mean bias of -1.39, whereas digit 1 was -2.61 and digit 2 was -3.84. Pearson correlations (r) for forehead, digit 1, and digit 2 with SO(2) were .834, .433, and .254, respectively. There were fewer unsuccessful measurements with the forehead oximetry technique.

CONCLUSIONS: Forehead sensors improve measurement of oxygen saturation in critically ill surgical/trauma patients at risk for decreased peripheral perfusion.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app