CASE REPORTS
JOURNAL ARTICLE
Add like
Add dislike
Add to saved papers

Significant caval penetration by the celect inferior vena cava filter: attributable to filter design?

This report describes transmural penetration of the inferior vena cava (IVC) by a newly introduced IVC filter within 9 days of its placement. A computed tomographic study demonstrated filter penetration with one of the primary struts lodging in the uncinate process of the pancreas. Because of the close resemblance of this new filter to another filter that has not been associated with penetration injuries, the key differences between the two designs were examined and the determination was made that the new filter, unlike the older one, has unprotected primary struts. Only filters with an unprotected primary strut design have been associated with penetration injuries such as the one described in this case.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app