Outcome of the Fidelis implantable cardioverter-defibrillator lead advisory: a report from the Canadian Heart Rhythm Society Device Advisory Committee.

BACKGROUND: The Medtronic Sprint Fidelis family of leads has recently been the subject of a widespread advisory. Lead failure rates are estimated at 2.3% at 30 months, 2.6 times the failure rate of the reference Medtronic 6947 lead.

OBJECTIVE: The purpose of this study was to contact pediatric and adult implantable cardioverter-defibrillator (ICD) implant centers across Canada to determine the short-term response to the October 15, 2007 Medtronic Fidelis lead advisory.

METHODS: All centers completed an 11-part survey to assess the frequency and presentation of lead failure, operator characteristics, and center's response.

RESULTS: Lead failure was noted in 80 (1.29%) of 6,181 patients at 21.0 months, with inappropriate shocks experienced in 45 (56%) of the 80 patients (overall risk 0.73%). No deaths were attributed to lead failure. Sensing was the primary form of failure, seen in 60 leads (75%), with pacing failure in 10 (13%), and high-voltage failure in 15 (19%). Assessment of the previous routine ICD interrogation prior to the advisory or lead failure demonstrated evidence of altered lead performance in only 8 (10%) of the 80 leads. Inappropriate shocks typically were multiple (median 7, range 1-122), with a single shock seen in only 5 patients. Lead failure was noted in 18 of 23 centers, representing 89.8% of leads implanted, with at least one failure noted in 15 of 16 centers that implanted more than 200 leads. Forty-seven of the 135 operators in the 23 institutions implanted the 80 leads that subsequently failed. Only 16 operators were involved in more than a single lead that subsequently failed; seven operators participated in three or more leads that subsequently failed. Seven centers planned to replace leads in most pacing-dependent patients, and two centers planned to replace leads in patients unable to hear the alert tone.

CONCLUSION: This national experience suggests a Fidelis lead failure rate of 1.29% at 21 months, most often presenting with multiple inappropriate shocks without evidence of impending failure from routine lead follow-up. Lead failure did not appear to cluster around specific operators or around high-volume or low-volume implant centers.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app