Failure of impedance monitoring to prevent adverse clinical events caused by fracture of a recalled high-voltage implantable cardioverter-defibrillator lead.

BACKGROUND: The Medtronic Sprint Fidelis implantable cardioverter-defibrillator high-voltage lead is prone to fracture. The October 2007 safety advisory recommended lead impedance monitoring to aid in identifying lead fractures.

OBJECTIVE: The aim of this single-center study was to examine the effectiveness of impedance monitoring for detecting Sprint Fidelis lead failures before they caused adverse clinical events such as inappropriate shocks.

METHODS: Impedance and sensing information were acquired during routine clinic and CareLink follow-up and at the time of lead failure using the Patient Alert, sensing integrity counter, nonsustained episode, and electrogram features in Medtronic pulse generators.

RESULTS: Between September 2004 and February 2008, 17 of 514 Sprint Fidelis leads (3.3%) followed up at our center failed between 11 and 35 months after implantation (mean 23.0 +/- 8.0 months). Fifteen of these failures (88%) were caused by pace-sense conductor fractures, and 2 (12%) were caused by high-voltage conductor defects. Twelve of 15 patients (80%) with pace-sense conductor fractures received inappropriate shocks; of these, 4 had no significant increase in lead impedance before they were shocked, 2 were shocked <3 hours after their lead impedances exceeded the 1,000 Omega audible alert threshold, and 2 patients did not hear the alarm. All pace-sense conductor failures whose sensing function could be evaluated (13 of 15) had oversensing based on stored data, and oversensing usually (11 of 13) preceded impedances changes.

CONCLUSION: Impedance monitoring did not prevent inappropriate shocks in two-thirds of our patients. Thus, pace-sense conductor impedance monitoring as currently implemented does not reliably forewarn patients of a lead malfunction. Consequently, patients who have Sprint Fidelis leads remain at risk for adverse clinical events associated with pace-sense conductor fracture.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app