The accuracy and efficacy of real-time continuous glucose monitoring sensor in patients with type 1 diabetes.
Diabetes Technology & Therapeutics 2008 October
BACKGROUND: The accuracy and efficacy of the Medtronic Diabetes (Northridge, CA) Real-Time (RT)-Continuous Glucose Monitoring (CGM) sensor were analyzed in 72 subjects with type 1 diabetes.
METHODS: This was a retrospective analysis of 60,050 temporally paired data points (sensor and glucose meter values) obtained during the course of an outpatient ambulatory study evaluating the efficacy of a sensor-augmented pump system in adults and adolescents. Subjects uploaded sensor values and self-monitoring blood glucose data to the CareLink Clinical Application (Medtronic Diabetes) via the Internet, every 2 weeks during the course of the study.
RESULTS: The overall percentage of sensor readings within +/-20% or +/-30% agreement of reference glucose readings was 75.6% and 86.8%, respectively. The highest rate of agreement occurred in the 240-400 mg/dL range, where 79.9% of sensor readings were within +/-20% of meter values and 91.5% of sensor readings were within 30% of meter values. The mean absolute relative difference for all subjects was 15.8%, and the median absolute relative difference was 10.9%. The bias was -2.13 mg/dL. Paired glucose measurements from the RT-CGM and meter demonstrated that 95.9% of paired points in the overall subject population fell in zones A and B of the Clarke Error Grid. Consensus Error Grid Analysis established that 99.2% of paired data points were in zones A and B.
CONCLUSIONS: This study reports the accuracy of a continuous glucose sensor with a large number of paired data points (60,050). RT-CGM is safe and well tolerated and provides readings that are in close agreement with glucose meter values.
METHODS: This was a retrospective analysis of 60,050 temporally paired data points (sensor and glucose meter values) obtained during the course of an outpatient ambulatory study evaluating the efficacy of a sensor-augmented pump system in adults and adolescents. Subjects uploaded sensor values and self-monitoring blood glucose data to the CareLink Clinical Application (Medtronic Diabetes) via the Internet, every 2 weeks during the course of the study.
RESULTS: The overall percentage of sensor readings within +/-20% or +/-30% agreement of reference glucose readings was 75.6% and 86.8%, respectively. The highest rate of agreement occurred in the 240-400 mg/dL range, where 79.9% of sensor readings were within +/-20% of meter values and 91.5% of sensor readings were within 30% of meter values. The mean absolute relative difference for all subjects was 15.8%, and the median absolute relative difference was 10.9%. The bias was -2.13 mg/dL. Paired glucose measurements from the RT-CGM and meter demonstrated that 95.9% of paired points in the overall subject population fell in zones A and B of the Clarke Error Grid. Consensus Error Grid Analysis established that 99.2% of paired data points were in zones A and B.
CONCLUSIONS: This study reports the accuracy of a continuous glucose sensor with a large number of paired data points (60,050). RT-CGM is safe and well tolerated and provides readings that are in close agreement with glucose meter values.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app