Add like
Add dislike
Add to saved papers

Percutaneous closure of perimembranous ventricular septal defects with Amplatzer occluders--a single centre experience.

Kardiologia Polska 2008 September
BACKGROUND: Perimembranous ventricular septal defect (VSD) is the most common congenital heart defect. Percutaneous transcatheter closure of VSD is one of the greatest challenges in interventional cardiology.

AIM: Presentation of our experience in transcatheter closure of perimembranous VSD.

METHODS: Eighteen patients were treated. Nine patients (group I) had VSD closed with implant Amplatzer Perimembranous VSD Occluder (PMVSDO) whereas the other nine had VSD closed with Muscular VSD Occluder (MVSDO). In the second group the presence of at least 4 mm rim from aortic valve was mandatory to undergo the procedure. Average patients age was 17.1 (3.2-40) years, defect diameter--4.7 (4-8) mm and Qp/Qs ratio--1.84 (1.5-4.6). Perimembranous interventricular septum aneurysm was noted in 5 cases. Only patients who had hemodynamically important defect (Qp/Qs ratio >1.5) were selected for interventional VSD closure. Patients with subarterial VSDs, pulmonary hypertension or/and aortic regurgitation were excluded. Ventricular septal defect closure was performed with standard techniques.

RESULTS: Procedures were completed successfully in 16 of 18 patients. There was no early or late implant embolisation. After the procedure in every case complete closure or important reduction of the shunt was observed. In the group I there was a trend towards more frequent occurrence of rhythm disturbances (p=0,08), including two cases with severe arrhythmias occurring during VSD closure requiring abandoning of procedure. In other 2 cases (patients age 12 and 14 years) in the second week after PMVSDO placement complete atrio-ventricular block occurred. In one patient sinus rhythm was restored after steroid treatment whereas another patient required pacemaker implantation. In group II mild nonprogressive tricuspid regurgitation was noted in 3 patients.

CONCLUSIONS: Percutaneous perimembranous VSD closure is an interesting alternative to surgical treatment. In selected cases closure of the defect with muscular VSD implant is effective and safe.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app