Journal Article
Research Support, Non-U.S. Gov't
Add like
Add dislike
Add to saved papers

Diagnostic performance of MR imaging in the assessment of subchondral fractures in avascular necrosis of the femoral head.

OBJECTIVE: A prospective study was conducted to determine the accuracy of routine magnetic resonance (MR) imaging in correctly identifying subchondral fracture in avascular necrosis of the femoral head without apparent focal collapse on standard radiographs.

MATERIALS AND METHODS: Spiral computed tomography (CT) with coronal and sagittal reformations and routine MR imaging with spin-echo T1WI and fat-suppressed spin-echo T2WI coronal, axial, and sagittal images were performed in 28 hips of 25 patients (M/F = 20:5; age 16-76 years) suffering from early-stage avascular necrosis of the femoral head on standard radiographs. The MR images were reviewed by a musculoskeletal radiologist and a general radiologist in blinded fashion. Using CT as the standard of reference, the accuracy of MR imaging in diagnosing subchondral fractures in avascular necrosis was evaluated.

RESULTS: When the diagnoses of the two readers were compared with each other, only 16 of the 28 diagnoses (57.5%) agreed. Seventeen of the 28 MR imaging readings (60.7%) made by the musculoskeletal radiologist and 15 of the 28 (53.5%) made by the general radiologist agreed with those of the CT standard. False-positive diagnosis (that is, diagnosis of fracture when no fracture could be seen on CT) was more common than false-negative diagnosis.

CONCLUSIONS: The accuracy of routine MR imaging in the evaluation of subchondral fracture is not satisfactory. False-positive diagnosis is not uncommon. Interpretation of routine MR imaging readout should be guarded.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app