Use of rapid influenza diagnostic tests under field conditions as a screening tool during an outbreak of the 2009 novel influenza virus: practical considerations.
Journal of Clinical Virology 2010 March
BACKGROUND: Rapid influenza diagnostic tests (RIDTs) are used in various settings as a first-line screen of patient specimens. During the initial outbreak of the 2009 novel influenza A/H1N1 virus, the Nebraska Public Health Laboratory (NPHL) adopted a testing algorithm, attempting to maximize the usefulness of RIDTs. However, it became apparent that a high percentage of the positive specimens received from off-site facilities were negative for influenza viruses by the confirmatory test, the Luminex xTAG Respiratory Viral Panel (RVP) molecular assay.
OBJECTIVES: To explore the cause of discrepancies between RIDTs results obtained from on-site facility testing versus confirmatory testing performed at NPHL.
STUDY DESIGN: Specimens (n=336) tested with RIDTs at off-site facilities and screened for high-probability of containing H1N1 were sent to the NPHL for confirmatory testing by RVP.
RESULTS: Of 336 specimens analyzed, 104 were negative for influenza A or B by both RIDT and RVP; 127 were positive by both tests; 102 were positive by RIDT only; and 3 were positive by RVP only. Using the RVP assay as the gold standard, overall RIDT characteristics in this screened population were: sensitivity=97.7% (95%CI: 92.5, 99.3); specificity=48.1% (95%CI: 40.4, 55.8); positive predictive value=54.3% (95%CI: 47.0, 61.4); and negative predicative value=97.1% (95%CI: 90.6, 99.1).
CONCLUSIONS: The results show that the confirmation of RIDT-positive results varied widely by testing site. Possible explanations for the discrepancies in performance characteristics include testing a narrowly defined sample population, test facility characteristics, facility work load, and seasonal timing.
OBJECTIVES: To explore the cause of discrepancies between RIDTs results obtained from on-site facility testing versus confirmatory testing performed at NPHL.
STUDY DESIGN: Specimens (n=336) tested with RIDTs at off-site facilities and screened for high-probability of containing H1N1 were sent to the NPHL for confirmatory testing by RVP.
RESULTS: Of 336 specimens analyzed, 104 were negative for influenza A or B by both RIDT and RVP; 127 were positive by both tests; 102 were positive by RIDT only; and 3 were positive by RVP only. Using the RVP assay as the gold standard, overall RIDT characteristics in this screened population were: sensitivity=97.7% (95%CI: 92.5, 99.3); specificity=48.1% (95%CI: 40.4, 55.8); positive predictive value=54.3% (95%CI: 47.0, 61.4); and negative predicative value=97.1% (95%CI: 90.6, 99.1).
CONCLUSIONS: The results show that the confirmation of RIDT-positive results varied widely by testing site. Possible explanations for the discrepancies in performance characteristics include testing a narrowly defined sample population, test facility characteristics, facility work load, and seasonal timing.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app