Accuracy of non-contact infrared thermometry versus rectal thermometry in young children evaluated in the emergency department for fever.
OBJECTIVE: We evaluated the accuracy of a non-contact infrared thermometer compared with a rectal thermometer.
METHODS: Two hundred patients, ages 1 month to 4 years, were included in the study. Each child underwent contemporaneous standard rectal thermometry and mid forehead non-contact infrared thermometry. Clinical features, including chief complaint, recently administered antipyretic agents, and ambient temperature at the time of measurement, were included.
ANALYSIS: Linear models were used to compare agreement between the 2 techniques, as well as to determine bias of infrared thermometry at different rectal temperatures. Multivariate linear models were used to evaluate the impact of clinical variables and ambient temperature.
RESULTS: A linear relationship between rectal and infrared temperature measurements was observed; however, the coefficient of determination (r(2)) value between was only 0.48 (P < 0.01). Infrared thermometry tended to overestimate the temperature of afebrile children and underestimate the temperature of febrile patients (P < .01). Ambient temperature and child age did not affect the accuracy of the device.
CONCLUSION: In this study, non-contact infrared thermometry did not sufficiently agree with rectal thermometer to indicate its routine use.
METHODS: Two hundred patients, ages 1 month to 4 years, were included in the study. Each child underwent contemporaneous standard rectal thermometry and mid forehead non-contact infrared thermometry. Clinical features, including chief complaint, recently administered antipyretic agents, and ambient temperature at the time of measurement, were included.
ANALYSIS: Linear models were used to compare agreement between the 2 techniques, as well as to determine bias of infrared thermometry at different rectal temperatures. Multivariate linear models were used to evaluate the impact of clinical variables and ambient temperature.
RESULTS: A linear relationship between rectal and infrared temperature measurements was observed; however, the coefficient of determination (r(2)) value between was only 0.48 (P < 0.01). Infrared thermometry tended to overestimate the temperature of afebrile children and underestimate the temperature of febrile patients (P < .01). Ambient temperature and child age did not affect the accuracy of the device.
CONCLUSION: In this study, non-contact infrared thermometry did not sufficiently agree with rectal thermometer to indicate its routine use.
Full text links
Related Resources
Trending Papers
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
British Society for Rheumatology guideline on management of adult and juvenile onset Sjögren disease.Rheumatology 2024 April 17
Albumin: a comprehensive review and practical guideline for clinical use.European Journal of Clinical Pharmacology 2024 April 13
Renin-Angiotensin-Aldosterone System: From History to Practice of a Secular Topic.International Journal of Molecular Sciences 2024 April 5
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app