Multicenter experience with extraction of the Sprint Fidelis implantable cardioverter-defibrillator lead.

OBJECTIVES: This study was undertaken to determine the safety and feasibility of extraction of the Sprint Fidelis (Medtronic, Minneapolis, Minnesota) lead.

BACKGROUND: The reported failure rate of the Sprint Fidelis defibrillator lead has increased to a range greater than initially appreciated with emerging evidence of an accelerating rate of fracture. At present, consensus guidelines continue to recommend against prophylactic extraction of the lead, citing major complication rates between 1.4% and 7.3%. However, data regarding the safety and feasibility of extraction of small-diameter, backfilled implantable cardioverter-defibrillator leads such as the Sprint Fidelis are limited.

METHODS: We performed a retrospective cohort study of consecutive patients undergoing extraction of Sprint Fidelis (models 6930, 6931, 6948, 6949) leads at 5 high-volume centers. Patient characteristics, indications for extraction, and use of countertraction sheath (CTS) assistance are reported. The risk of major and minor complications was determined. A multivariable logistic regression model was developed to predict factors associated with the use of CTS assistance.

RESULTS: Between May 2005 and August 2009, 349 Sprint Fidelis leads were extracted from 348 patients. All leads were removed completely. The average duration of the implanted lead was 27.5 months (range 0.03 to 58.8 months). Approximately one-half of the extracted leads were fractured (49.4%), and 26.5% were extracted prophylactically. The other major indication for extraction was infection (22.8%). Extraction was achieved with simple traction in 49.4% leads; CTS assistance was required in 174 cases (50.6%). In multivariable models, length of time since implantation was directly related to the need for CTS assistance (odds ratio per month since implantation: 1.035; 95% confidence interval: 1.010 to 1.061; p=0.006). There were no major procedural complications or deaths.

CONCLUSIONS: Extraction of the Sprint Fidelis lead can be performed safely by experienced operators at high-volume centers with a complication rate lower than that reported for older generation leads. However, leads with longer implant durations are associated with the use of CTS assistance. Recommendations regarding prophylactic Sprint Fidelis lead extraction may warrant reconsideration.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app