Clinical utility of Abbott Precision Xceed Pro® ketone meter in diabetic patients.
Pediatric Diabetes 2011 November
BACKGROUND: Diagnosis and management of diabetic ketoacidosis (DKA) often rely on the measurement of urine ketones along with blood glucose, anion gap, and pH. These values, however, do not reliably reflect the severity of ketoacidosis. The Abbott Precision Xceed Pro® meter is an FDA-approved device that quantitatively measures β-hydroxybutyrate (BOH) in whole blood. This study was undertaken to determine whether the ketone meter meets the analytical criteria to aid DKA diagnosis and management in the hospital.
METHODS: 54 heparinized venous whole blood BOH concentrations from 27 diabetic patients were measured by the Abbott meter, and compared with the plasma BOH concentrations measured with Stanbio reagent (reference method). Measurements were done in the hospital central laboratory.
RESULTS: Of the 54 pairs of specimens analyzed, 17 pairs displayed a difference of >15% between the two methods. Nearly all discrepant points occurred when BOH >5 mmol/L (reference method). Linearity evaluation revealed that the meter is not linear from 0.0 to 8.0 mmol/L, contrary to the claim by the manufacturer. Further, we identified acetoacetate, a metabolite commonly present in DKA patients, as a potential interfering substance for the meter BOH measurement.
CONCLUSIONS: BOH measurements by the Abbott meter up to 3 mmol/L correlate well with the reference method, but become discrepant above that point. While this characteristic may be useful in the diagnosis of DKA, it may not allow clinicians to serially follow the response to therapy in hospitalized DKA patients with BOH values greater than 5 mmol/L (reference method).
METHODS: 54 heparinized venous whole blood BOH concentrations from 27 diabetic patients were measured by the Abbott meter, and compared with the plasma BOH concentrations measured with Stanbio reagent (reference method). Measurements were done in the hospital central laboratory.
RESULTS: Of the 54 pairs of specimens analyzed, 17 pairs displayed a difference of >15% between the two methods. Nearly all discrepant points occurred when BOH >5 mmol/L (reference method). Linearity evaluation revealed that the meter is not linear from 0.0 to 8.0 mmol/L, contrary to the claim by the manufacturer. Further, we identified acetoacetate, a metabolite commonly present in DKA patients, as a potential interfering substance for the meter BOH measurement.
CONCLUSIONS: BOH measurements by the Abbott meter up to 3 mmol/L correlate well with the reference method, but become discrepant above that point. While this characteristic may be useful in the diagnosis of DKA, it may not allow clinicians to serially follow the response to therapy in hospitalized DKA patients with BOH values greater than 5 mmol/L (reference method).
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app