Posterior corneal curvature assessment after Epi-LASIK for myopia: comparison of Orbscan II and Pentacam imaging.
Korean Journal of Ophthalmology : KJO 2012 Februrary
PURPOSE: To compare the changes in posterior corneal curvature using scanning slit topography (Orbscan II) and Scheimpflug imaging (Pentacam) before and after Epi-laser in situ keratomileusis (LASIK) for myopia.
METHODS: In a prospective observational case-series study, 20 myopic patients having undergone Epi-LASIK were examined serially with two different devices, Orbscan II and Pentacam, preoperatively and one month postoperatively. Posterior central elevation (PCE) and posterior maximal elevation (PME) were compared between the two devices, and the changes in parameters after Epi-LASIK were analyzed using a difference map.
RESULTS: All parameters (preoperative and postoperative PCE and preoperative and postoperative PME) that were measured using the Orbscan II were significantly greater compared to those of the Pentacam (for all p < 0.001). PCE and PME were significantly increased one month postoperatively in the Orbscan II measurements (p < 0.05) but were not significantly increased in the Pentacam measurements. Also, ΔPCE and ΔPME, in the difference map obtained by each serial scanning, were significantly greater in the Orbscan II measurements than with the Pentacam (p = 0.012, p = 0.016).
CONCLUSIONS: The Pentacam measurements displayed significantly reduced values in all parameters related to posterior corneal elevation compared to those of the Orbscan II. The Pentacam showed no significant change in posterior corneal curvature after Epi-LASIK, based on the difference map.
METHODS: In a prospective observational case-series study, 20 myopic patients having undergone Epi-LASIK were examined serially with two different devices, Orbscan II and Pentacam, preoperatively and one month postoperatively. Posterior central elevation (PCE) and posterior maximal elevation (PME) were compared between the two devices, and the changes in parameters after Epi-LASIK were analyzed using a difference map.
RESULTS: All parameters (preoperative and postoperative PCE and preoperative and postoperative PME) that were measured using the Orbscan II were significantly greater compared to those of the Pentacam (for all p < 0.001). PCE and PME were significantly increased one month postoperatively in the Orbscan II measurements (p < 0.05) but were not significantly increased in the Pentacam measurements. Also, ΔPCE and ΔPME, in the difference map obtained by each serial scanning, were significantly greater in the Orbscan II measurements than with the Pentacam (p = 0.012, p = 0.016).
CONCLUSIONS: The Pentacam measurements displayed significantly reduced values in all parameters related to posterior corneal elevation compared to those of the Orbscan II. The Pentacam showed no significant change in posterior corneal curvature after Epi-LASIK, based on the difference map.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app