OBJECTIVE: To compare the detectability of simulated pulmonary nodules on chest radiographs between an irradiation side sampling indirect flat-panel detector (ISS-FPD) and computed radiography (CR).
MATERIALS AND METHODS: This study was an observer performance study. Simulated pulmonary nodules of 8 mm in diameter were superimposed on an anthropomorphic chest phantom. Chest radiographs were acquired under 2 exposure levels (4 and 3.2 mAs) with the ISS-FPD and the CR. Six thoracic radiologists evaluated all 40 images (10 patterns × 2 different exposure doses × 2 different systems) for the presence or absence of a lesion over each of 12 defined areas on a 3-megapixel monochrome liquid-crystal display. Receiver operating characteristic (ROC) curves were obtained for observation in predefined 480 areas. A jackknife method was used for statistical analysis. Differences with a P value of <0.05 were considered significant.
RESULTS: The analysis of the observer detection of simulated pulmonary nodules showed larger areas under the ROC curve (AUC) by the ISS-FPD than by the CR. There was a statistically significant difference between the two systems at 3.2 mAs (P=0.0330).
CONCLUSION: The ISS-FPD was superior to the CR for the detection of simulated pulmonary nodules at 3.2 mAs.
MATERIALS AND METHODS: This study was an observer performance study. Simulated pulmonary nodules of 8 mm in diameter were superimposed on an anthropomorphic chest phantom. Chest radiographs were acquired under 2 exposure levels (4 and 3.2 mAs) with the ISS-FPD and the CR. Six thoracic radiologists evaluated all 40 images (10 patterns × 2 different exposure doses × 2 different systems) for the presence or absence of a lesion over each of 12 defined areas on a 3-megapixel monochrome liquid-crystal display. Receiver operating characteristic (ROC) curves were obtained for observation in predefined 480 areas. A jackknife method was used for statistical analysis. Differences with a P value of <0.05 were considered significant.
RESULTS: The analysis of the observer detection of simulated pulmonary nodules showed larger areas under the ROC curve (AUC) by the ISS-FPD than by the CR. There was a statistically significant difference between the two systems at 3.2 mAs (P=0.0330).
CONCLUSION: The ISS-FPD was superior to the CR for the detection of simulated pulmonary nodules at 3.2 mAs.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app