We have located links that may give you full text access.
CASE REPORTS
JOURNAL ARTICLE
Collapse of the Advanta V12 Large Diameter covered stent following implantation for aortic coarctation.
Catheterization and Cardiovascular Interventions 2014 January 2
We report a case of collapse of the Advanta V12 Large Diameter (LD) covered stent following previous successful implantation for native aortic coarctation. This unreported complication was successfully managed with subsequent stent redilation and implantation of two covered Cheatham-Platinum stents within the collapsed Advanta V12 stent. This case highlights the importance of correct stent apposition to the aortic wall and also raises some concerns regarding the radial strength of the Advanta V12 stent.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app