A biomechanical study on proximal plate fixation techniques in periprosthetic femur fractures.
Injury 2014 January
INTRODUCTION: Proximal plate fixation is a crucial factor in osteosynthesis of periprosthetic femur fractures. Stability and strength of different fixation concepts for proximal plate fixation were compared.
MATERIALS AND METHODS: Twelve fresh frozen, bone mineral density matched human femora, instrumented with cemented hip endoprosthesis were osteotomized simulating a Vancouver B1 fracture. Specimens were instrumented with locking compression plates, fixed proximally with either locking attachment plate (LAP), monocortical screws, cerclage plus monocortical screws (1cerclage) or cerclages only (4cerclages). Cyclic testing was performed with monotonically increasing load until failure. Relative movements at the proximal plate-femur interface were registered by motion tracking.
RESULTS: The LAP construct exhibited a significantly longer cumulative survival (failure criterion 1mm separation at the proximal plate fixation) compared to the monocortical (p=0.048) and 4cerclages constructs (p=0.012) but not to 1cerclage constructs.
CONCLUSION: Bicortical screw anchorage improves proximal plate fixation in periprosthetic fractures. The cerclage-screw combination is a valuable alternative especially in osteoporotic bone.
MATERIALS AND METHODS: Twelve fresh frozen, bone mineral density matched human femora, instrumented with cemented hip endoprosthesis were osteotomized simulating a Vancouver B1 fracture. Specimens were instrumented with locking compression plates, fixed proximally with either locking attachment plate (LAP), monocortical screws, cerclage plus monocortical screws (1cerclage) or cerclages only (4cerclages). Cyclic testing was performed with monotonically increasing load until failure. Relative movements at the proximal plate-femur interface were registered by motion tracking.
RESULTS: The LAP construct exhibited a significantly longer cumulative survival (failure criterion 1mm separation at the proximal plate fixation) compared to the monocortical (p=0.048) and 4cerclages constructs (p=0.012) but not to 1cerclage constructs.
CONCLUSION: Bicortical screw anchorage improves proximal plate fixation in periprosthetic fractures. The cerclage-screw combination is a valuable alternative especially in osteoporotic bone.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app