Influence of maternal body mass index on accuracy and reliability of external fetal monitoring techniques.

OBJECTIVE: To evaluate the performance of external electronic fetal heart rate and uterine contraction monitoring according to maternal body mass index.

DESIGN: Secondary analysis of prospective equivalence study.

SETTING: Three US urban teaching hospitals.

SAMPLE: Seventy-four parturients with a normal term pregnancy.

METHODS: The parent study assessed performance of two methods of external fetal heart rate monitoring (abdominal fetal electrocardiogram and Doppler ultrasound) and of uterine contraction monitoring (electrohystero-graphy and tocodynamometry) compared with internal monitoring with fetal scalp electrode and intrauterine pressure transducer. Reliability of external techniques was assessed by the success rate and positive percent agreement with internal methods. Bland-Altman analysis determined accuracy. We analyzed data from that study according to maternal body mass index.

MAIN OUTCOME MEASURES: We assessed the relationship between body mass index and monitor performance with linear regression, using body mass index as the independent variable and measures of reliability and accuracy as dependent variables.

RESULTS: There was no significant association between maternal body mass index and any measure of reliability or accuracy for abdominal fetal electrocardiogram. By contrast, the overall positive percent agreement for Doppler ultrasound declined (p = 0.042), and the root mean square error from the Bland-Altman analysis increased in the first stage (p = 0.029) with increasing body mass index. Uterine contraction recordings from electrohysterography and tocodynamometry showed no significant deterioration related to maternal body mass index.

CONCLUSIONS: Accuracy and reliability of fetal heart rate monitoring using abdominal fetal electrocardiogram was unaffected by maternal obesity, whereas performance of ultrasound degraded directly with maternal size. Both electrohysterography and tocodynamometry were unperturbed by obesity.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app