EZ-IO(®) intraosseous device implementation in German Helicopter Emergency Medical Service.

Resuscitation 2015 March
BACKGROUND: Intraosseous access (IO) is a rapid and safe alternative when peripheral venous access is difficult. Our aim was to summarize the first three years experience with the use of a semi-automatic IO device (EZ-IO(®)) in German Helicopter Emergency Medical Service (HEMS).

METHODS: Included were all patients during study period (January 2009-December 2011) requiring an IO access performed by HEMS team. Outcome variables were IO rate, IO insertion success rates, site of IO access, type of EZ-IO(®) needle set used, strategy of vascular access, procedure related problems and operator's satisfaction.

RESULTS: IO rate was 0.3% (348/120.923). Overall success rate was 99.6% with a first attempt success rate of 85.9%; there was only one failure (0.4%). There were three insertion sites: proximal tibia (87.2%), distal tibia (7.5%) and proximal humerus (5.3%). Within total study group IO was predominantly the second-line strategy (39% vs. 61%, p<0.001), but in children<7 years, in trauma cases and in cardiac arrest IO was more often first-line strategy (64% vs. 28%, p<0.001; 48% vs. 34%, p<0.032; 50% vs. 29%, p<0.002 respectively). Patients with IO access were significantly younger (41.7±28.7 vs. 56.5±24.4 years; p<0.001), more often male (63.2% vs. 57.7%; p=0.037), included more trauma cases (37.3% vs. 30.0%; p=0.003) and more often patients with a NACA-Score≥5 rating (77.0% vs. 18.6%; p<0.001). Patients who required IO access generally presented with more severely compromised vital signs associated with the need for more invasive resuscitation actions such as intubation, chest drains, CPR and defibrillation. In 93% EZ-IO(®) needle set handling was rated "good". Problems were reported in 1.6% (needle dislocation 0.8%, needle bending 0.4% and parafusion 0.4%).

CONCLUSIONS: The IO route was generally used in the most critically ill of patients. Our relatively low rate of usage would indicate that this would be compatible with the recommendations of established guidelines. The EZ-IO(®) intraosseous device proved feasible with a high success rate in adult and pediatric emergency patients in HEMS.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app