Contrast-enhanced voiding urosonography phantom study: intravenous iodinated and gadolinium-based contrast agents may cause false-negative results in assessment of vesicoureteral reflux in children.
Pediatric Radiology 2015 June
BACKGROUND: Contrast-enhanced voiding urosonography (ce-VUS) is commonly requested simultaneously to other diagnostic imaging necessitating intravenous contrast agents. To date there is limited knowldedge about intravesical interactions between different types of contrast agents.
OBJECTIVE: To assess the effect of excreted intravenous iodinated and gadolinium-based contrast agents on the intravesical distribution of ultrasound contrast within contrast-enhanced voiding urosonography.
MATERIALS AND METHODS: Iodinated (iomeprol, iopamidol) and gadolinium-based (gadoterate meglumine) contrast agents were diluted to bladder concentration and injected into balloons filled with saline solution. CT scans were performed to assess the contrast distribution in these phantoms. Regions of interest were placed at the top and bottom side of each balloon and Hounsfield units (HU) were measured. Three other balloons were filled with saline solution and contrast media likewise. The ultrasound contrast agent sulphur hexafluoride was added and its distribution was assessed using sonography.
RESULTS: MDCT scans showed a separation of two liquid layers in all bladder phantoms with the contrast layers located at the bottom and the saline solution at the top. Significant differences of the HU measurements at the top and bottom side were observed (P < 0.001-0.007). Following injection of ultrasound contrast agent, US showed its distribution exclusively among the saline solution.
CONCLUSIONS: False-negative results of contrast-enhanced voiding urosonography may occur if it is performed shortly after imaging procedures requiring intravenous contrast.
OBJECTIVE: To assess the effect of excreted intravenous iodinated and gadolinium-based contrast agents on the intravesical distribution of ultrasound contrast within contrast-enhanced voiding urosonography.
MATERIALS AND METHODS: Iodinated (iomeprol, iopamidol) and gadolinium-based (gadoterate meglumine) contrast agents were diluted to bladder concentration and injected into balloons filled with saline solution. CT scans were performed to assess the contrast distribution in these phantoms. Regions of interest were placed at the top and bottom side of each balloon and Hounsfield units (HU) were measured. Three other balloons were filled with saline solution and contrast media likewise. The ultrasound contrast agent sulphur hexafluoride was added and its distribution was assessed using sonography.
RESULTS: MDCT scans showed a separation of two liquid layers in all bladder phantoms with the contrast layers located at the bottom and the saline solution at the top. Significant differences of the HU measurements at the top and bottom side were observed (P < 0.001-0.007). Following injection of ultrasound contrast agent, US showed its distribution exclusively among the saline solution.
CONCLUSIONS: False-negative results of contrast-enhanced voiding urosonography may occur if it is performed shortly after imaging procedures requiring intravenous contrast.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app