We have located links that may give you full text access.
Evaluation Study
Journal Article
Safe balloon sizing for endoscopic dilatation of subglottic stenosis in children.
Journal of Laryngology and Otology 2017 March
OBJECTIVES: To describe our experience and provide guidelines for maximum safe balloon sizes according to age in children undergoing balloon dilatation.
METHOD: A retrospective review was conducted of children undergoing balloon dilatation for subglottic stenosis in a paediatric tertiary unit between May 2006 and February 2016.
RESULTS: A total of 166 patients underwent balloon dilatation. Mean ( ± standard deviation) patient age was 4.5 ± 3.99 years. The median balloon size was 8 mm, the median balloon inflation pressure was 10 atm, and the mean balloon inflation time was 65.1 ± 18.6 seconds. No significant unexpected events occurred. The Pearson correlation co-efficient for the relationship between patient age and balloon size was 0.85 (p = 0.001), suggesting a strongly positive correlation.
CONCLUSION: This study demonstrated that balloon dilatation is a safe procedure for airway stenosis. The results suggest using a balloon diameter that is equal to the outer diameter of the age-appropriate endotracheal tube +1 mm for the larynx and subglottis and +2 mm for the trachea.
METHOD: A retrospective review was conducted of children undergoing balloon dilatation for subglottic stenosis in a paediatric tertiary unit between May 2006 and February 2016.
RESULTS: A total of 166 patients underwent balloon dilatation. Mean ( ± standard deviation) patient age was 4.5 ± 3.99 years. The median balloon size was 8 mm, the median balloon inflation pressure was 10 atm, and the mean balloon inflation time was 65.1 ± 18.6 seconds. No significant unexpected events occurred. The Pearson correlation co-efficient for the relationship between patient age and balloon size was 0.85 (p = 0.001), suggesting a strongly positive correlation.
CONCLUSION: This study demonstrated that balloon dilatation is a safe procedure for airway stenosis. The results suggest using a balloon diameter that is equal to the outer diameter of the age-appropriate endotracheal tube +1 mm for the larynx and subglottis and +2 mm for the trachea.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app