Ceramic Bearings with Titanium Adapter Sleeves Implanted During Revision Hip Arthroplasty Show Minimal Fretting or Corrosion: a Retrieval Analysis.
BACKGROUND: The BIOLOX® option system, consisting of a BIOLOX® delta ceramic femoral head with a titanium alloy adapter sleeve, is being increasingly utilized in revision hip arthroplasty. The sleeve protects the ceramic head from fracture and improper motion about the stem trunnion when a damaged trunnion is encountered at revision surgery. Corrosion and fretting due to metal-metal contact at the taper region of hip prosthesis create the potential of causing periprosthetic osteolysis and adverse local tissue reactions.
QUESTIONS/PURPOSES: The objective of this study was to identify the type and extent of damage to retrieved sleeves and ceramic heads to determine their in vivo performance.
METHODS: Twenty-four ceramic heads with titanium alloy sleeves were examined. The articular and taper surfaces for each ceramic head were assessed for metal transfer using a subjective grading system. All surfaces of the 24 titanium sleeves and stem trunnions (only available for 7 of 24 cases) were assessed for corrosion and fretting using an established grading system. Scanning electron microscopy and energy dispersive X-ray analysis were conducted on representative sample of sleeves.
RESULTS: Fretting and corrosion were higher at the inner surface of the taper sleeve than the outer sleeve. Mean fretting scores at the inner taper and outer taper sleeve surfaces were 1.8 and 1.2, respectively. The mean corrosion score at the inner taper surface was 1.8; no corrosion was observed on the outer surface of any taper sleeve. SEM and EDS analyses provided further indications of low levels of damage.
CONCLUSION: Fretting and corrosion were less severe than previously reported for conventional THA metal-metal taper connections, indicating that a ceramic head and titanium sleeve is a safe alternative in revision THA.
QUESTIONS/PURPOSES: The objective of this study was to identify the type and extent of damage to retrieved sleeves and ceramic heads to determine their in vivo performance.
METHODS: Twenty-four ceramic heads with titanium alloy sleeves were examined. The articular and taper surfaces for each ceramic head were assessed for metal transfer using a subjective grading system. All surfaces of the 24 titanium sleeves and stem trunnions (only available for 7 of 24 cases) were assessed for corrosion and fretting using an established grading system. Scanning electron microscopy and energy dispersive X-ray analysis were conducted on representative sample of sleeves.
RESULTS: Fretting and corrosion were higher at the inner surface of the taper sleeve than the outer sleeve. Mean fretting scores at the inner taper and outer taper sleeve surfaces were 1.8 and 1.2, respectively. The mean corrosion score at the inner taper surface was 1.8; no corrosion was observed on the outer surface of any taper sleeve. SEM and EDS analyses provided further indications of low levels of damage.
CONCLUSION: Fretting and corrosion were less severe than previously reported for conventional THA metal-metal taper connections, indicating that a ceramic head and titanium sleeve is a safe alternative in revision THA.
Full text links
Related Resources
Trending Papers
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Prevention and treatment of ischaemic and haemorrhagic stroke in people with diabetes mellitus: a focus on glucose control and comorbidities.Diabetologia 2024 April 17
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Eosinophilic Esophagitis: Clinical Pearls for Primary Care Providers and Gastroenterologists.Mayo Clinic Proceedings 2024 April
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app