Temporary Circulatory Support in U.S. Children Awaiting Heart Transplantation.

BACKGROUND: Extracorporeal membrane oxygenation (ECMO) has long served as the standard of care for short-term mechanical circulatory support in pediatrics. It is unknown whether newer-generation temporary circulatory support (TCS) devices afford children a meaningful survival advantage over ECMO.

OBJECTIVES: This study sought to determine whether bridge-to-heart transplant survival with a TCS device is superior to ECMO after adjusting for patient differences.

METHODS: All children ≤21 years of age listed for heart transplant from 2011 to 2015 who received a TCS device or ECMO as a bridge to transplant were identified using Organ Procurement and Transplantation Network data. Children supported with a TCS device were compared with a propensity score (PS)-matched cohort of children supported with ECMO as a bridge to transplant. The primary endpoint was Kaplan-Meier survival to transplant.

RESULTS: The number of TCS devices implanted in children increased from ≤3 per year before 2011 to 50 in 2015. Overall, 93 patients implanted with TCS devices were included for analysis (59% left ventricular assist devices, 23% right ventricular assist devices, 18% biventricular assist devices). The most commonly used device was the CentriMag-PediMag system (65%), followed by TandemHeart (18%), Rotaflow (6%), and Impella (5%). Among 164 PS-matched patients, support duration was longer for the TCS cohort (median 19 days vs. 6 days; p < 0.001), and was longest for the CentriMag-PediMag (24 days vs. 6 days; p < 0.001) with 27% supported for >60 days. Compared with the ECMO cohort, the PS-matched TCS cohort had longer survival to transplant (hazard ratio: 0.49; 95% confidence interval: 0.30 to 0.79) and longer overall survival (hazard ratio: 0.61; 95% confidence interval: 0.39 to 0.96), with 90-day mortality before transplant that was modestly reduced (from 45% with ECMO to 39% with TCS).

CONCLUSIONS: The use of TCS devices in children as a bridge to transplant has risen rapidly in recent years, led by the growth of magnetically levitated centrifugal flow pumps. Compared with conventional ECMO, TCS durations are longer, and more importantly, patient survival is superior.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app