Three-year Outcomes of a Randomized Controlled Trial Comparing a 4.5-mm-Wide to a 3.75-mm-Wide Titanium Implant for Bone Conduction Hearing.
Otology & Neurotology 2018 June
OBJECTIVE: To compare 3-year implant stability, survival, and tolerability of a 4.5-mm-wide (test) and a 3.75-mm-wide (control) percutaneous titanium implant for bone-conduction hearing, loaded with the sound processor after 3 weeks.
METHODS: Sixty implants were allocated in a 2:1 ratio (test-control) in 57 adult patients included in this prospective randomized controlled clinical trial. Follow-up visits were performed at 7, 14, 21, and 28 days; 6 and 12 weeks; 6 months; and at 1, 2, and 3 years after implantation. During these visits, the implant stability quotient (ISQ) was measured by means of resonance frequency analysis (RFA). The peri-abutment soft tissue status was assessed according to the Holgers classification. Skin height around the abutment was evaluated.
RESULTS: The mean area-under-the-curve (AUC) of ISQ-low was statistically significantly higher for the test implant (65.7 versus 61.4, p = 0.0002). Both implants showed high survival rates (97.4% versus 95.0%, p = 0.6374). Adverse soft tissue reactions were observed sporadically, with no significant inter-group differences. Skin thickening was seen in the majority of the patients, but no correlation with adverse soft tissue reactions or implant type was observed.
CONCLUSION: The 4.5-mm-wide implant provides significantly higher ISQ values during the first 3 years after surgery compared with the previous generation 3.75-mm-wide implant. Both implants showed high survival rates and good tolerability. These long-term results indicate that the wider implant, loaded with a sound processor at 3 weeks, is a safe and well-performing option for hearing rehabilitation in specific types of hearing loss.
METHODS: Sixty implants were allocated in a 2:1 ratio (test-control) in 57 adult patients included in this prospective randomized controlled clinical trial. Follow-up visits were performed at 7, 14, 21, and 28 days; 6 and 12 weeks; 6 months; and at 1, 2, and 3 years after implantation. During these visits, the implant stability quotient (ISQ) was measured by means of resonance frequency analysis (RFA). The peri-abutment soft tissue status was assessed according to the Holgers classification. Skin height around the abutment was evaluated.
RESULTS: The mean area-under-the-curve (AUC) of ISQ-low was statistically significantly higher for the test implant (65.7 versus 61.4, p = 0.0002). Both implants showed high survival rates (97.4% versus 95.0%, p = 0.6374). Adverse soft tissue reactions were observed sporadically, with no significant inter-group differences. Skin thickening was seen in the majority of the patients, but no correlation with adverse soft tissue reactions or implant type was observed.
CONCLUSION: The 4.5-mm-wide implant provides significantly higher ISQ values during the first 3 years after surgery compared with the previous generation 3.75-mm-wide implant. Both implants showed high survival rates and good tolerability. These long-term results indicate that the wider implant, loaded with a sound processor at 3 weeks, is a safe and well-performing option for hearing rehabilitation in specific types of hearing loss.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app