We have located links that may give you full text access.
The effects of gel-based manicure on pulse oximetry.
Singapore Medical Journal 2019 August
INTRODUCTION: Pulse oximetry is the standard monitoring technique of functional oxygen saturation (SpO2 ). As the use of fingernail polish has been described to alter SpO2 readings, its removal is commonly recommended prior to measurement. Gel-based manicures have gained popularity in recent years due to their attractiveness and longevity. However, the removal of gel nail polish requires a specialised procedure. Valuable time and resources can be saved if removal can be avoided. To our knowledge, there are no available studies on the effect of gel-based manicures on pulse oximetry readings. Hence, we evaluated the effect with two oximeters, using different technology and wavelength combinations.
METHODS: 17 healthy female adult volunteers were recruited for this single-blind randomised controlled trial. Subjects with hypothermia, hypotension, poor plethysmographic waveform and nail pathology were excluded. Colours tested were: black, purple, navy blue, green, light blue, white, yellow, orange, pink and red. Pulse oximetry was measured at 15- and 30-second intervals using two different pulse oximeters, the Philips M1191BL and Masimo SET®. Means were compared using paired t-tests.
RESULTS: Using the Masimo oximeter, light blue (ΔM = 0.97% ± 0.96%; p = 0.001) and orange (ΔM = 0.76 ± 1.17%; p = 0.016) gel nail polish resulted in a statistically significant increase from baseline SpO2 readings. With the Philips oximeter, the limits of agreement ranged from 2% for pink to 17% for black, indicating imprecision.
CONCLUSION: Gel-based manicures can result in overestimations of actual readings, delaying detection of hypoxaemia. Gel nail polish should be routinely removed or an alternative monitoring technique sought.
METHODS: 17 healthy female adult volunteers were recruited for this single-blind randomised controlled trial. Subjects with hypothermia, hypotension, poor plethysmographic waveform and nail pathology were excluded. Colours tested were: black, purple, navy blue, green, light blue, white, yellow, orange, pink and red. Pulse oximetry was measured at 15- and 30-second intervals using two different pulse oximeters, the Philips M1191BL and Masimo SET®. Means were compared using paired t-tests.
RESULTS: Using the Masimo oximeter, light blue (ΔM = 0.97% ± 0.96%; p = 0.001) and orange (ΔM = 0.76 ± 1.17%; p = 0.016) gel nail polish resulted in a statistically significant increase from baseline SpO2 readings. With the Philips oximeter, the limits of agreement ranged from 2% for pink to 17% for black, indicating imprecision.
CONCLUSION: Gel-based manicures can result in overestimations of actual readings, delaying detection of hypoxaemia. Gel nail polish should be routinely removed or an alternative monitoring technique sought.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app