Safety and efficiency of low-field magnetic resonance imaging in patients with cardiac rhythm management devices.

PURPOSE: Low-field magnetic resonance imaging (MRI), i.e. MRI with a static magnetic field strength <0.5 T, has been reported to be safe in patients with pacemakers, however there are no data about the safety of low-field MRI in patients with implantable cardioverter defibrillators (ICD) and/or cardiac resynchronization therapy (CRT). We aimed to investigate the safety and diagnostic efficiency of routine low-field MRI in patients with different devices for cardiac rhythm management (i.e. pacemakers and ICD, including devices with CRT).

METHOD: MRI scans of 446 regions of interest were evaluated with field strength of 0.2 T in 338 patients (62% male; age at MRI scan 76.1 ± 9.2 years; time since device implantation 4.1 ± 3.2 years) with cardiac rhythm management devices (298 pacemakers, 25 ICD, 8 CRT-ICD, and 7 CRT pacemakers). This analysis included 62 pacemaker-dependent patients (18.3%), 52 patients with 1.5-Tesla-MR conditional pacemakers (15.4%) and 13 patients with abandoned leads (3.9%).

RESULTS: Except for one examination, which was interrupted because of recurrent severe nausea, all MRI scans could be analyzed efficiently. No induction of arrhythmia or inhibition of pacemaker function occurred. Compared to the device interrogation before MRI, there were no significant changes in battery voltage, pacing capture threshold, sensing of intrinsic ECG, lead impedance, as well as shock impedance in ICD devices after completed examination.

CONCLUSIONS: Low-field MRI examinations (0.2 T) were efficient and safe regarding clinical and technical complications in patients with devices for cardiac rhythm management, even in case of pacemaker-dependency or the presence of abandoned leads.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app