Further observations on the effect of balloon size on the short term and intermediate term results of balloon dilatation of the pulmonary valve.
British Heart Journal 1988 December
The relation between the size of the balloon used for dilatation of the pulmonary valve and the extent of relief of pulmonary stenosis both immediately after and at intermediate term follow up was studied. Sixty four balloon dilatation procedures in 56 patients were divided into group 1-12 in which the ratio of the diameter of the balloon to that of the pulmonary annulus was less than or equal to 1.0--and group 2-52 in which the ratio was greater than 1.0. Both groups had similar mean (SD) pressure drops across the pulmonary valve before dilatation. Immediately after dilatation there was a significant reduction in the pulmonary valve gradient in both group 1 (84.3 (39.2) v 43.6 (26.8) mm Hg) and group 2 (92.8 (42.1) v 22.4 (13.6)mm Hg). On intermediate term follow up (6-34 months), residual pulmonary valve gradients were significantly lower in group 2 (20.8 (18.5)mm Hg) than in group 1 (75.0 (49.4) mm Hg), suggesting that restenosis was more common after dilatation with small balloons. These data suggest that although the immediate results with either small or large balloons are good, balloons that are larger than the pulmonary valve annulus produce more sustained relief of pulmonary stenosis. Further analysis within the group treated with larger balloons showed that the subgroup with a balloon/annulus ratio of 1.01-1.2 had more recurrences of stenosis (need for repeat balloon dilatation and larger number of patients with residual pulmonary valve gradients greater than 30 mm Hg) than subgroups with balloon/annulus ratios for diameters of 1.21-1.41, greater than 1.41, and greater than 1.5, in which there were no recurrences. Balloons that were > 1.5 times the size of the pulmonary valve annulus had no additional advantage over the other subgroups, namely, 1.21-- 1.4 and > 1.41. These results and reports of damage to the right ventricular outflow tract by oversized (> 1.5) balloons indicate that balloons that give a balloon/annulus ratio 1.2 -- 1.5 are the best for dilatation of the pulmonary valve.
Full text links
Related Resources
Trending Papers
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Anti-Arrhythmic Effects of Heart Failure Guideline-Directed Medical Therapy and Their Role in the Prevention of Sudden Cardiac Death: From Beta-Blockers to Sodium-Glucose Cotransporter 2 Inhibitors and Beyond.Journal of Clinical Medicine 2024 Februrary 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app